Dose distribution in a radiosurgery treatment for an acoustic neuroma

Authors

DOI:

https://doi.org/10.15392/bjrs.v9i2.1654

Keywords:

Radiosurgery, acoustic neuroma, radiochromic film

Abstract

Evaluation of the dose distribution delivered in a radiosurgery treatment of an acoustic neuroma using an adapted Alderson phantom and radiochromic film to assess absolute dose. A system was developed to perform the dose measurements, using the Alderson Rando phantom´s head carrying a modified slice, where a material reproducing the tumor was inserted. The phantom’s head with the modified slice was irradiated in a medical linear accelerator, using a procedure similar to that adopted in real cases, and radiochromic films were used to assess dose values. The suitability of the dose distribution delivered in the treatment was evaluated by comparing the Dose Volume Histograms, obtained from the Treatment Planning System (TPS) and by radiochromic film measurements. An agreement around 3% between experimental data and TPS´s calculations, showing an acceptable concordance with the planning results. The established approach of transforming 2D arrays of dose in a 3D one is satisfactory, validating it. The developed method shows to be an excellent alternative to quality control in radiosurgery using the radiochromic film to assess absolute dose.

Downloads

Download data is not yet available.

References

Leksell L. Sterotactic method and radiosurgery of the brain. Acta Chir Scand 102:316–319, 1951.

Mascarenhas F, Costa M, Ortiz M, Almeida A, Carvalho H, Ferreira A, Cattoni M. Radiocirurgia Estereotáxica em Tumores Benignos e Malignos do Sistema Nervoso Central. Acta Med Port 18:45-60, 2005.

TECDOC 1151. Aspectos Físicos da Garantia da Qualidade em Radioterapia – Protocolo de Controle da Qualidade, Ministério da Saúde, INCa – Instituto Nacional de Câncer, Rio de Janeiro, RJ, Brasil, 2001.

Garcia-Garduno OA, Celis MA, Lárraga-Gutiérrez JM, Moreno-Jiménez S, Mattínes-Dávalos A, Rodr´´iguez-Villafuerte M. Radiation transmission, leakage and beam penumbra measurements of micro-multileaf collimator using GafChromic EBT film. J. Appl. Clin Med Phys 9:2802, 2008.

Moutsatsos M, Petrokokkinos L, Zourari K, Papagiannis P, Karaiskos P, Dardoufas K, Damila-his J, Seimenis I, Georgiou E. Gamma Knife relative dosimetry using VIP polymer gel and EBT radiochromic films. Journal of Physics 164: 012053, 2009.

Powell C, Micallef C, Gonsalves A, Wharram B, Ashley S, Brada M. Fractioned stereotactic radiotherapy in the treatment of vestibular schwannoma (acoustic neuroma): predicting the risk of hydrocephalus. Int. J. Radiation Oncology Biol. Phys. 80: 11433–1150, 2011.

Ferreira T, Rasband W. 4 August 2014. ImageJ User Guide: imagej.nih.gov/ij/docs/guide/user-guide.pdf.

Alves VGL, Cardoso SC, da Silva AX. Gafchromic EBT2 dosimetry via robust optimization. Computer Physics Communications 184: 1708-1716, 2013.

Bouchard H, Lacroix F, Beaudoin G, Carrier JF, Kawrakow I. On the characterization and uncertainty analysis of radiochromic film dosimetry. Med. Phys 36: 1931-1946, 2009.

Devic S. Radiochromic film dosimetry: Past, present, and future. Physica Medica 27: 122-134, 2011.

FilmQA Pro Home Page. Film QA Pro. Web. 25 March 2012. http://www.filmqapro.com/FilmQA_Pro.htm