Radiology and mammography standard X-ray spectra simulated with the Monte Carlo method

Authors

  • Joel Marques Xavier Filho
  • Iury Santos Silveira
  • Linda Viola Ehlin Caldas

DOI:

https://doi.org/10.15392/bjrs.v9i2C.1670

Keywords:

EGSnrc, TRS-457, HVL

Abstract

Six standard beams described in the TRS-457 (IAEA): RQR 5, 8, M1, M2, M3, M4 were simulated using the EGSnrc Monte Carlo code. Each spectrum was created by an X-ray tube simulated in BEAMnrc, and attenuation curves were obtained using the application egs_kerma. The quality of each beam was evaluated by the 1st and 2nd half-value layers, the homogeneity coefficients and the mean energies. All beams presented quality parameters compatible with those described in TRS-457 (IAEA).

Downloads

Download data is not yet available.

References

IAEA - International Atomic Energy Agency. Dosimetry in diagnostic radiology: an international code of practice. Technical Reports Series No. 457. Vienna, 2007.

NAZÁRIO, M.; KHOURY, H. J.; HAZIN, C. Calibração de equipamentos de radioproteção com radiação gama no Laboratório de Metrologia das Radiações Ionizantes - DEN/UFPE. Metrologia 2003. Pernambuco, 2003.

PERINI, A. P.; NEVES, L. P.; SANTOS, W. S.; CALDAS, L. V. E. Experimental and Monte Carlo evaluation of an ionization chamber in a 60Co beam. Journal of Physics: Conference Series 733 012081, 2016.

HERRNSDORF, L.; ANDERSSON, M. ; GUNNARSSON, M. ; MATTSSON, S. Optimization of energy and directional response of a small 4π silicon dosemeter for quality control of CT/CBCT-units – A 3D CAD, Monte-Carlo, AM approach. Applied Radiation and Isotopes, v. 158 – 109050, p. 1 -12, 2020.

SALEHI, Z. ; BALVASI, E. ; AZIZ, M. Z. A. A review of the recent Monte Carlo (MC) simulation for dosimetry in mammographic applications. Malaysian Journal of Medicine and Health Sciences, v. 14(SUPP1), p. 76-81, 2018.

IAEA - International Atomic Energy Agency. Calibration of radiation protection monitoring instruments. Safety Reports Series No. 16. Vienna, 2000.

KAWRAKOW, I.; ROGERS, D. W. O. The EGSnrc code system: Monte Carlo simulation of electron and photon transport. National Research Council of Canada, NRCC. Report PIRS-701, 2003.

ROGERS, D. W. O.; WALTERS, B.; KAWRAKOW, I. BEAMnrc User’s Manual. National Research Council of Canada, NRCC. Report PIRS-0509(A)revL, 2010.

KAWRAKOW, I.; MAINEGRA-HING, E.; TESSIER, F.; TOWNSON, R.; WALTERS, B. EGSnrc C++ class library. Report PIRS-898, 2019.

MA, C.-M. and ROGERS, D. W. O. BEAMDP User’s Manual. National Research Council of Canada, NRCC. Report PIRS-0509(E)revA, 2020.

YU, M. K.; SLOBODA, R. S.; MURRAY, B. Linear accelerator photon beam quality at off-axis points. Medical Physics, v. 24(2), p. 233–239, 1997.

MATHIEU, K. B.; KAPPADATH S. C.; WHITE R. A.; ATKINSON E. N.; CODY D. D. An empirical model of diagnostic x-ray attenuation under narrow-beam geometry. Medical Physics, v. 38(8), p. 4546-4555, 2011.

ARNS, A. M. Towards clinical implementation of ultrafast combined kV-MV cone-beam CT for IGRT of lung tumors within breath-hold: evaluation of dosimetry and registration accuracy based on phantom studies. Doctorate dissertation – Heidelberg University, Heidelberg, 2016.

BAUR, M.; UHLMANN, N.; PÖSCHEL, T.; SCHRÖTER, M. Correction of beam hardening in X-ray radiograms. Review of Scientific Instruments, v. 90, 025108, 2019.

Downloads

Published

2021-08-08

Issue

Section

Congresso Brasileiro de Metrologia das Radiações Ionizantes - CBMRI 2020

How to Cite

Radiology and mammography standard X-ray spectra simulated with the Monte Carlo method. Brazilian Journal of Radiation Sciences, Rio de Janeiro, Brazil, v. 9, n. 2C (Suppl.), 2021. DOI: 10.15392/bjrs.v9i2C.1670. Disponível em: https://bjrs.org.br/revista/index.php/REVISTA/article/view/1670.. Acesso em: 1 may. 2024.

Similar Articles

1-10 of 29

You may also start an advanced similarity search for this article.