Image quality assessment using NEMA NU 4/2008 standards in small animal PET scanner
DOI:
https://doi.org/10.15392/bjrs.v7i2A.580Palabras clave:
Image Quality, Positron Emission Tomography, NEMA.Resumen
In Brazil, there are few microPET in use and a quality control protocols standardization are needed to harmonize their use in the research field. Thus, the purpose of this study is to characterize the image quality performance of the microPET scanner (LabPET 4, GE healthcare Technologies, Waukesha, WI) using the NEMA NU 4/ 2008 standards and specific phantom. The NEMA image-quality (IQ) phantom consists of 3 different regions to analyze distinct characteristics: image noise (%SD), expressed as percentage SD in a uniform region (%SD), recovery coefficient (RC) and Spill-over (SOR) in air and water. The IQ phantom was filled with 18F-FDG calibrated at the beginning of acquisition, placed in the center of the field-of-view (FOV) and measured with the typical whole body imaging protocol. The images were reconstructed with different reconstruction methods (FBP-2D; MLEM-3D and OSEM-3D); with and without high resolution (HR) when possible. The results were compared. The LabPET 4 system produces appropriate image and with performance according to the literature. The present study is an initial step to verify the NEMA NU 4/2008 use in the Brazilian scenario for further standardization.
Descargas
Referencias
YAO R.; LECOMTE R.; CRAWFORD E. Small-Animal PET: What is it, and why do we need it? Journal of Nuclear Medicine Technology. 2011. Vol. 40 nº3, pp.157-165.
NEMA - National Electrical Manufacturers Association. Performance Measurements of Small Animal Positron Emission Tomographs. Rosslyn VA; 2008 Standards Publication NU 4-2008.
GE Healthcare Technologies, Triumph Service Guide Technical Publication. Revision Draft 6, Copyright. Available at: <http://www.gehealthcare.com>. 2011.
TRETRAULT M. A., et al. System architecture of the LabPET small animal PET scanner. IEEE Transactions on Nuclear Science, 2008. Vol.55, pp. 2546-2550.
FONTAINE R., et al. The hardware and signal processing architecture of LabPETTM, a small animal APD-based digital PET scanner. IEEE Transactions on Nuclear Science, February 2009. Vol. 56, pp. 3-9.
DISSELHORST J. A., et al. Image-Quality Assessment for several Positron Emitters using the NEMA NU 4-2008 Standards in the Siemens Inveon Small-Animal PET Scanner. The Journal of Nuclear Medicine. April 2010. Vol. 51 nº4, pp. 610-617.
ZAIDI H. and RATIB O. NEMA NU-04-based Performance Characteristics of the LabPET-8TM Small Animal PET Scanner. 2001. IEEE Nuclear Science Symposium Conference Record. pp. 3654-3660.
Descargas
Publicado
Número
Sección
Licencia
Derechos de autor 2021 Brazilian Journal of Radiation Sciences (BJRS)

Esta obra está bajo una licencia internacional Creative Commons Atribución 4.0.
Licencia: los artículos de BJRS tienen una licencia internacional Creative Commons Attribution 4.0, que permite el uso, el intercambio, la adaptación, la distribución y la reproducción en cualquier medio o formato, siempre que se otorgue el crédito correspondiente al autor o autores originales y a la fuente, proporcione un enlace a la licencia Creative Commons e indique si se realizaron cambios. Las imágenes u otros materiales de terceros en el artículo están incluidos en la licencia Creative Commons del artículo, a menos que se indique lo contrario en una línea de crédito al material. Si el material no está incluido en la licencia Creative Commons del artículo y su uso previsto no está permitido por la regulación legal o excede el uso permitido, el autor deberá obtener el permiso directamente del titular de los derechos de autor. Para ver una copia de esta licencia, visite http://creativecommons.org/licenses/by/4.0/