Diagnostic x-ray dosimeters using standard float zone (FZ) and XRA-50 commercial diodes
DOI:
https://doi.org/10.15392/bjrs.v7i2A.592Palabras clave:
X-ray dosimetry, Si photodiodes dosimeters, Solid-state dosimetryResumen
The results obtained with a standard float zone (FZ) silicon diode, processed at the Helsinki Institute of Physics, used as on-line diagnostic X-ray dosimeter are described in this work. The device was connected in the short-circuit current mode to the input of an integrating electrometer. The response repeatability and the current sensitivity coefficient of the diode were measured with diagnostic X-ray beams in the range of 40-80 kV. The dose-response of the device, evaluated from 10 mGy up to 500 mGy, was linear with high charge sensitivity. Nevertheless, significant energy dependence was observed in the charge sensitivity of FZ device for energies below 70 kV. The dosimetric characteristics of this FZ diode were compared to those of an XRA-50 commercial Si diode, specially designed to X-ray dosimetry. The results obtained with the FZ diode evidenced that it can be an alternative choice for diagnostic X-ray dosimetry, although it needs to be calibrated for individual X-ray beam energies. The studies of long-term stability and the radiation hardness of these diodes are under way.
Descargas
Referencias
KHOURY, H. J.; HAZIN, C. A.; MASCARENHAS, A. P.; da SILVA Jr., E. F. Low cost silicon photodiode for electron dosimetry. Radiation Protection Dosimetry, v. 84, p. 341-343, 1999.
GRUSELL, E. and MEDIN, J. General characteristics of the use of silicon diode detectors for clinical dosimetry in proton beams. Phys. Med. Biol., v. 45, p. 2573-2582, 2000.
CASATI, M.; BRUZZI, M.; BUCCIOLINI, M.; MENICHELLI, D.; SCARINGELLA, M.; PIEMONTE, C.; FRETWURST, E. Characterization of standard and oxygenated float zone Si diodes under radiotherapy beams. Nucl. Instrum. Methods Phys. Res., Sect. A, v. 552, p. 158-162, 2005.
LINDSTROM, G.; MOLL, M.; FRETWURST, E. Radiation hardness of silicon detectors – a challenge from high-energy physics. Nucl. Instrum. Methods Phys. Res., Sect. A, v. 426, p. 1-15, 1999.
CANDELORI, A. Radiation-hard detectors for very high luminosity colliders. Nucl. Instrum. Methods Phys. Res., Sect. A, v. 560, p. 103-107, 2006.
BOSISIO, L.; DITTONGO, S.; QUAI, E.; RACHEVSKAIA, I. Observation of substrate-type inversion in high-resistivity silicon structures irradiated with high-energy electrons. IEEE Trans. Nucl. Sci., v. 50, p. 219-22, 2003.
LI, Z.; LI, C. J.; VERBITSKAYA, E. Study of bulk damage in high resistivity silicon detectors irradiated by high dose of Co-60 gamma-radiation. IEEE Trans. Nucl. Sci., v. 44, p. 834-839, 1997.
HARKÖNEN, J.; ABERU, M.; ANBINDERIS, P.; ANBINDERIS, T.; D’AMBROSIO, N.; DE BOER, W., et.al. Recent results from the CERN RD39 collaboration on super-radiation hard cryogenic silicon detectors for LHC and LHC upgrade. Nucl. Instrum. Methods Phys. Res., Sect. A., v. 535, p. 384-388, 2004.
CAMARGO, F.; KHOURY, H. J.; NASCIMENTO, C. R.; ASFORA, V. K.; BUENO, C. C. Evaluation of a multi-guard ring (MGR) structure diode as diagnostic X-ray dosimeter. Nucl. Instrum. Methods Phys. Res., Sect. A, v. 580, p. 194-196, 2007.
CAMARGO, F.; GONÇALVES, J. A. C.; KHOURY, H. J.; NAPOLITANO, C. M.; HARKONEN, J.; BUENO, C. C. MCz diode response as a high-dose gamma radiation dosimeter. Radiat. Meas., v. 43, p. 1160-1162, 2008.
KHOURY, H. J.; SCHELIN, H.; SOBOLL, D.; LUNELLI, N.; BAPTISTA, C. Evaluation of commercial silicon diode for electron dosimetry. Nucl. Instrum. Methods Phys. Res., Sect. A, v. 580, p. 537-539, 2007.
Descargas
Publicado
Número
Sección
Licencia
Derechos de autor 2021 Brazilian Journal of Radiation Sciences (BJRS)

Esta obra está bajo una licencia internacional Creative Commons Atribución 4.0.
Licencia: los artículos de BJRS tienen una licencia internacional Creative Commons Attribution 4.0, que permite el uso, el intercambio, la adaptación, la distribución y la reproducción en cualquier medio o formato, siempre que se otorgue el crédito correspondiente al autor o autores originales y a la fuente, proporcione un enlace a la licencia Creative Commons e indique si se realizaron cambios. Las imágenes u otros materiales de terceros en el artículo están incluidos en la licencia Creative Commons del artículo, a menos que se indique lo contrario en una línea de crédito al material. Si el material no está incluido en la licencia Creative Commons del artículo y su uso previsto no está permitido por la regulación legal o excede el uso permitido, el autor deberá obtener el permiso directamente del titular de los derechos de autor. Para ver una copia de esta licencia, visite http://creativecommons.org/licenses/by/4.0/