Portable grazing exit X-ray fluorescence system using a low-power X-ray tube
DOI:
https://doi.org/10.15392/bjrs.v7i2A.714Palavras-chave:
GE-XRF, Portable system, Trace elements analysisResumo
In this work was developed a portable system of grazing exit X-ray fluorescence (geometric 90° - 0°) that can be applied in several areas science and technology. GE-XRF portable system is formed by a mini X-ray tube of low power (anode of Au) and a SiPIN detector. The reflectors used as sample support (sampler carrier) were quartz discs. The grazing exit angle was experimentally determined by measuring a cooper solution (10 μg.g-1). The accuracy of the system was checked using multielement reference solution as standard reference material. The relative errors between measured and certified values are in the range of 4 to 19%. The first results showed a background was drastically reduced at grazing exit angles, enabling trace elemental analysis. The system of GE-XRF proved to be quite stable and reproducible. This paper shows that it is possible to produce a portable system of grazing exit X-ray fluorescence compact, efficient, low-cost and easy-to-handle instrumentation using a low power X-ray tube and a SiPIN compact detector.Downloads
Referências
Klockenkämper, R.; von Bohlen A. Total-Reflection X-Ray Fluorescence Analysis and Relat-ed Methods, 2nd ed. New Jersey & United States: John Wiley & Sons, 2015.
Ashida T.; Tsuji, K. Development of a compact grazing exit X-ray fluorescence spectrometer for fast trace elemental analysis. Spectrochimica Acta Part B, v. 101, pp. 200–203, 2014.
Nowak, S. H.; Banaś, D.; Błchucki, W.; Cao, W.; Dousse, J.-Cl.; Hönicke, P.; Hoszowska, J.; Jabłoński, Ł.; Kayser, Y.; Kubala-Kukuś, A.; Pajek, M.; Reinhardt, F.; Savu, A. V.; Szlachetko, J. Grazing angle X-ray fluorescence from periodic structures on silicon and silica surfaces. Spectro-chimica Acta Part B, v. 98, pp. 65-75, 2014.
Becker, R. S.; Golovchenko, J. A.; Patel, J. R. X-ray evanescent-wave absorption and emission. Physical Review Letters, v. 50, pp. 153–156, 1983.
Urbach, H. P.; de Bokx, P. K. Calculation of intensities in grazing-emission X-ray fluorescence, Physical Review B, v. 53, pp. 3752-3763, 1996.
Kayser, Y.; Sá, J.; Szlachetko, J. Depth-Resolved X‑ray Absorption Spectroscopy by Means of Grazing Emission X‑ray Fluorescence, Analytical Chemistry, v. 87, pp. 10815−10821, 2015.
Tsuji, K.; Takenaka, H.; Wagatsuma, K.; de Bokx, P. K.; Van Grieken, R. E. Enhancement of X-ray fluorescence intensity from an ultra-thin sandwiched layer at grazing-emission angles. Spectro-chimica Acta Part B: Atomic Spectroscopy, v. 54, pp. 1881-1888, 1999.
Emoto, T.; Sato, Y.; Konishi, Y.; Ding, X.; Tsuji, K. Development and applications of grazing exit micro X-ray fluorescence instrument using a polycapillary X-ray lens. Spectrochimica Acta Part B, v. 59, pp. 1291– 1294, 2004.
Yang, J.; Tsuji, K.; Lin, X.; Han, D.; Ding, X.; A micro X-ray fluorescence analysis method us-ing polycapillary X-ray optics and grazing exit geometry. Thin Solid Films, v. 517, pp. 3357–3361, 2009.
Awane, T.; Fukuoka, S.; Nakamachi, K.; Tsuji, K. Grazing Exit Micro X-ray Fluorescence Analysis of a Hazardous Metal Attached to a Plant Leaf Surface Using an X-ray Absorber Method. Analytical Chemistry, v. 81, pp. 3356–3364, 2009.
Kayser, Y.; Szlachetko, J.; Banaś, D.; Cao, W.; Dousse, J.-Cl.; Hoszowska, J.; Kubala-Kukuś, A.; Pajek, M. High-energy-resolution grazing emission X-ray fluorescence applied to the characteri-zation of thin Al films on Si. Spectrochimica Acta Part B, v. 88, pp. 136–149, 2013.
Solé, V. A.; Papillon, E.; Cotte, M.; Walter, Ph.; Susini, J. A multiplatform code for the analysis of energy-dispersive X-ray fluorescence spectra, Spectrochimica Acta Part B, v. 62, pp. 63-68, 2007.
Curie, L. A. Limits for quantitative detection and quantitative determination. Analytical Chemistry, v. 40, pp. 586-593, 1968.
Tiwari, M. K.; Singh, A. K.; Sawhney, K. J. S. Sample Preparation for Evaluation of Detection Limits in X-ray Fluorescence Spectrometry. Analytical Sciences, v. 21, pp. 143-147, 2005.
Downloads
Publicado
Edição
Seção
Categorias
Licença
Direitos autorais (c) 2021 Brazilian Journal of Radiation Sciences

Este trabalho está licenciado sob uma licença Creative Commons Attribution 4.0 International License.
Declaro que o presente artigo é original, não tendo sido submetido à publicação em qualquer outro periódico nacional ou internacional, quer seja em parte ou em sua totalidade. Declaro, ainda, que uma vez publicado na revista Brazilian Journal of Radiation Sciences, editada pela Sociedade Brasileira de Proteção Radiológica, o mesmo jamais será submetido por mim ou por qualquer um dos demais co-autores a qualquer outro periódico. Através deste instrumento, em meu nome e em nome dos demais co-autores, porventura existentes, cedo os direitos autorais do referido artigo à Sociedade Brasileira de Proteção Radiológica, que está autorizada a publicá-lo em meio impresso, digital, ou outro existente, sem retribuição financeira para os autores.
Licença
Os artigos do BJRS são licenciados sob uma Creative Commons Atribuição 4.0 Licença Internacional, que permite o uso, compartilhamento, adaptação, distribuição e reprodução em qualquer meio ou formato, desde que você dê o devido crédito ao (s) autor (es) original (is) e à fonte, forneça um link para a licença Creative Commons, e indique se mudanças foram feitas. As imagens ou outro material de terceiros neste artigo estão incluídos na licença Creative Commons do artigo, a menos que indicado de outra forma em uma linha de crédito para o material. Se o material não estiver incluído no licença Creative Commons do artigo e seu uso pretendido não é permitido por regulamentação legal ou excede o uso permitido, você precisará obter permissão diretamente do detentor dos direitos autorais. Para visualizar uma cópia desta licença, visite http://creativecommons.org/licenses/by/4.0/