Comparative Study of the Use of Bismuth Shielding on Head Computed Tomography Scans Using Phantoms
DOI:
https://doi.org/10.15392/bjrs.v8i1B.936Palavras-chave:
Computed Tomography, Bismuth Shielding, DosimetryResumo
Computed Tomography (CT) has been one of the most used exam for radiologic diagnostic in medicine. The increase of CT is a global concern due to high doses of radiation. The dose evaluation in CT is one of many steps that can contribute for reducing patient doses. The head CT scans helps to diagnose disorders that affect the brain, including tumors, infarction, bleeding within the brain, hematoma and other diseases. The aim of this work is to compare the reduction of absorbed dose in the head CT scan with and without the use of bismuth shielding using a male and female anthropomorphic phantom model Alderson Rando.The head CT scan were done from the cervical vertebra C1 to the top of skull, in a GE CT scanner, LightSpeed VCT model, with 64 channels.Radiochromic films strips were used to evaluate the doses in the organs such as lenses, thyroid, hypophysis, spinal cord, pharynx, breasts, salivary and parotid glands.Record doses were lower with the use of bismuth shielding in both phantoms for all organs, mainly in the lenses. The analysis of noise in the image of the head central slice presented acceptable values for soft tissues, less than 1%. Dose values were significantly reduced and they suggested that the use of bismuth shielding would be a proper procedure for protection during a head CT scan.Downloads
Referências
ALONSO, T.C; MOURÃO, A.P; SANTANA, P.C; SILVA, T.A. Assessment of breast absorbed doses during thoracic computed tomography scan to evaluate the effectiveness of bismuth shielding. Applied Radiation and Isotopes, v. 117, p. 55-57, 2016.
INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION ICRP, Statement on tis-sue reactions/early and late effects of radiation in normal tissues and organs-threshold doses for tissue reactions in a radiation protection context, Publication 118 (2012).
P. Colleti et al., “To shield or not to shield: application of bismuth breast shields”, AJR, Vol. 200, pp. 503-507 (2013).
GIADDUI, T; CUI, Y. J; GALVIN, W; CHEN, Y; YU, Y; XIAO, Y. Characteristics of Gafchromic XRQA2 films for kV image dose measurement. Medical Physics, v. 39, p. 842-850, 2012.
MOURÃO, A. P; ALONSO, T. C; SILVA, T. A. Dose profile variation with voltage in head CT scans using radiochromic films. Radiation Physics and Chemistry, v. 95, p. 254-257, 2014.
COSTA, K. C; GOMEZ, A. M. L; ALONSO, T. C; MOURÃO, A. P. Radiochromic film calibration for the RQT9 quality beam. Radiantion Physics and Chemistry, v. 140, p. 370-372, 2017.
Downloads
Publicado
Edição
Seção
Categorias
Licença
Direitos autorais (c) 2020 Brazilian Journal of Radiation Sciences

Este trabalho está licenciado sob uma licença Creative Commons Attribution 4.0 International License.
Declaro que o presente artigo é original, não tendo sido submetido à publicação em qualquer outro periódico nacional ou internacional, quer seja em parte ou em sua totalidade. Declaro, ainda, que uma vez publicado na revista Brazilian Journal of Radiation Sciences, editada pela Sociedade Brasileira de Proteção Radiológica, o mesmo jamais será submetido por mim ou por qualquer um dos demais co-autores a qualquer outro periódico. Através deste instrumento, em meu nome e em nome dos demais co-autores, porventura existentes, cedo os direitos autorais do referido artigo à Sociedade Brasileira de Proteção Radiológica, que está autorizada a publicá-lo em meio impresso, digital, ou outro existente, sem retribuição financeira para os autores.
Licença
Os artigos do BJRS são licenciados sob uma Creative Commons Atribuição 4.0 Licença Internacional, que permite o uso, compartilhamento, adaptação, distribuição e reprodução em qualquer meio ou formato, desde que você dê o devido crédito ao (s) autor (es) original (is) e à fonte, forneça um link para a licença Creative Commons, e indique se mudanças foram feitas. As imagens ou outro material de terceiros neste artigo estão incluídos na licença Creative Commons do artigo, a menos que indicado de outra forma em uma linha de crédito para o material. Se o material não estiver incluído no licença Creative Commons do artigo e seu uso pretendido não é permitido por regulamentação legal ou excede o uso permitido, você precisará obter permissão diretamente do detentor dos direitos autorais. Para visualizar uma cópia desta licença, visite http://creativecommons.org/licenses/by/4.0/