Comparative evaluation of gamma cameras performance for internal monitoring of workers exposed to I-131 in nuclear medicine services
DOI:
https://doi.org/10.15392/bjrs.v8i1A.996Palavras-chave:
internal monitoring, nuclear medicine, radiation protection, iodine 131Resumo
Introduction: In the nuclear medicine practice, a variety of radionuclides are handled for diagnostic and therapy which represents a significant risk of internal and external exposure for workers. Such occupational exposure requires the implementation of a routine monitoring plan, including intakes of radionuclides. Currently, in Brazil, approximately 90 Nuclear Medicine Services (NMS) are authorized by the Nacional Nuclear Energy Commission of Brazil to handle 131I for therapy purposes [1], resulting in a significant number of workers routinely exposed to internal exposures. 131I remains as one of the most frequently used radionuclides in NMS, and presents higher risks of intake and internal doses. The International Atomic Energy Agency recommends implementing an internal monitoring program of this group of workers. However, in Brazil there are not qualified laboratories to attend the demand of internal monitoring [2].
The IRD has developed several studies proposing the use of detectors available in NMS for 131I occupational monitoring. Among such detectors, the gamma camera stands out to present the highest sensitivities for the proposed application. This work presents a comparison of the performances of gamma cameras available in six NMS in Brazil for 131I in vivo monitoring purpose.
Methodology: The technique consists basically in (i) determining the calibration factors for the measurement of 131I in the thyroid, (ii) estimation of the minimum detectable activities and the corresponding Minimum detectable intakes and effective doses. The IRD neck-thyroid phantom containing a 133Ba source certified by the LNMRI of IRD was used to calibrate the medical devices evaluated in this study.
The standard geometry consisted in positioning the phantom at 12 cm distance to the Gamma Cameras (GC) detectors. The count time was determined according to detectors sensitivity for a measurement performed at 1 and 7 days after the intake, considering a weekly generic monitoring frequency, resulting in 48 monitoring periods per year.
Calibration of the Detection System
The Calibration Factor (CF) is calculated as follows:
CFctg/Bq = (total counts – total background counts) / 131I Eq Ac
Where: 131I Eq Ac = Equivalent activity of 131I in neck-thyroid phantom, corrected for the calibration date.
Evaluation of the Sensitivity of the Methods
The Minimum Detectable Activity (MDA) [3] is calculated as follows:
MDABq = (4,65 x √N) / CF
Where: N = Total background counts ; and CF = Calibration factor of the technique.
The Minimum Detectable Intake (MDI) is calculated as follows:
MDIBq = AMD / m(t).
Where: MDA = Minimum Detectable Activity; and m(t) = Retention fraction of 131I at time t.
The Minimum Detectable Effective Dose (MDED) is calculated as follows:
MDEDmSv = IMD x e(g).
Where: MDI = Minimum detectable intake; and e(g) = Dose coefficient in mSv/Bq intake of 131I by inhalation.
In order to be considered useful for internal dosimetry purposes, the technique should, at least, be able to detect an activity that would result in an annual effective dose below 1 mSv for the most likely internal exposure scenario [5].
The monitoring period is determinant to evaluate the sensitivity of the proposed method to be applied in routine internal monitoring in a given NMS. This evaluation can be made by the following equation:
Annual Sensitivity mSv = MDED x n.
Where: MDED = Minimum detectable effective dose; and n = Total number of measurements in 1 year.
Results: Seven gamma cameras of six different NMS were included in this study. Table 1 presents the results for MDA and Annual Sensitivity for each equipment calibrated in those facilities.
Table 1: MDA and Annual Sensitivities of the detectors analyzed for 48 monitoring periods per year.
Manufacturer
Model
Facility
Count Time
MDA (Bq)
Annual Sensitivity (mSv)
1 day
7 days
Phillips
BrightView XCT
A
3
130
0.5
0.9
GE
Millennium
B
7
134
0.6
0.9
GE
Discovery MN/CT 670
C
3
146
0.6
1.0
Siemens
e.cam 180
D
3
151
0.6
1.0
Elscint
Apex SPX-6
E
3
146
0.6
1.0
Siemens
Symbia
E
3
129
0.5
0.9
GE
Millennium
F
2
131
0.5
0.9
In order to avoid disturbing the routine of the NMS, the count time was optimized to be as low as possible but keeping the necessary sensitivity of the method for the application in occupational monitoring. Furthermore the monitoring plan must be established according to the schedule of the radionuclide handling of the facility.
Conclusions: The gamma camera performance can be evaluated by analyzing the MDA of the detection system as well as the count time required for the system to reach the minimum sensitivity to be considered useful for internal monitoring purpose. This study evaluated modern and older gamma camera models, and despite the difference of year of manufacture between the models, all gamma cameras presented enough sensitivity, considering a generic weekly monitoring frequency, for use as a monitoring device up to 7 days after intake. In most cases, the appropriate sensitivity was achieved setting the count time to 3 minutes, which can be considered a feasible and short count time. It shows that in addition to being simple, the implementation of the proposed methodology is easy and inexpensive.
Downloads
Referências
CNEN - Comissão Nacional de Energia Nuclear. Instalações Autorizadas. Available at: < http://www.cnen.gov.br/instalacoes-autorizadas>. Last accessed: 28 Apr. 2019.
DANTAS B. M.; CARDOSO J. S.; DANTAS A. L. A.; et al. Intercomparação Nacional de Medição In Vivo de I-131 na Tireoide – Projeto TC IAEA BRA 9055. Scientia Plena, v. 9, 2013.
HPS - Health Physics Society. Performance Criteria for Radiobioassay, N13.30, 1996.
IAEA - International Atomic Energy Agency. Methods for Assessing Occupational Radiation Doses due to Intakes of Radionuclides. Safety Report Series No. 37, Vienna, 2004.
IAEA - International Atomic Energy Agency. Methods for Assessing Occupational Radiation Doses due to Intakes of Radionuclides. Safety Report Series No. 37, Vienna, 2004.
Downloads
Publicado
Edição
Seção
Categorias
Licença
Direitos autorais (c) 2020 Brazilian Journal of Radiation Sciences

Este trabalho está licenciado sob uma licença Creative Commons Attribution 4.0 International License.
Declaro que o presente artigo é original, não tendo sido submetido à publicação em qualquer outro periódico nacional ou internacional, quer seja em parte ou em sua totalidade. Declaro, ainda, que uma vez publicado na revista Brazilian Journal of Radiation Sciences, editada pela Sociedade Brasileira de Proteção Radiológica, o mesmo jamais será submetido por mim ou por qualquer um dos demais co-autores a qualquer outro periódico. Através deste instrumento, em meu nome e em nome dos demais co-autores, porventura existentes, cedo os direitos autorais do referido artigo à Sociedade Brasileira de Proteção Radiológica, que está autorizada a publicá-lo em meio impresso, digital, ou outro existente, sem retribuição financeira para os autores.
Licença
Os artigos do BJRS são licenciados sob uma Creative Commons Atribuição 4.0 Licença Internacional, que permite o uso, compartilhamento, adaptação, distribuição e reprodução em qualquer meio ou formato, desde que você dê o devido crédito ao (s) autor (es) original (is) e à fonte, forneça um link para a licença Creative Commons, e indique se mudanças foram feitas. As imagens ou outro material de terceiros neste artigo estão incluídos na licença Creative Commons do artigo, a menos que indicado de outra forma em uma linha de crédito para o material. Se o material não estiver incluído no licença Creative Commons do artigo e seu uso pretendido não é permitido por regulamentação legal ou excede o uso permitido, você precisará obter permissão diretamente do detentor dos direitos autorais. Para visualizar uma cópia desta licença, visite http://creativecommons.org/licenses/by/4.0/