Influence of the resolution mode on mean and maximum SUV for PET images acquired by a LabPET SOLO 4 scanner
DOI:
https://doi.org/10.15392/2319-0612.2023.2273Palabras clave:
LabPET SOLO 4, Resolution Mode, Standardized Uptake ValueResumen
In preclinical PET, SUV (standardized uptake value) is a robust quantifier that can be used to analyze PET images. Several factors – biological or technical – can affect SUV determination. Among technical factors, it is possible to cite the reconstruction protocols of PET images. This work evaluated the influence of two resolution modes – standard and high – on mean and maximum SUVs. The PET image of a mouse with a tumor in left flank was chosen from a PET image bank and reconstructed using two different protocols varying the resolution mode. The post-processing of images was performed using AMIDE software and eight volumes of interest (VOIs) were defined. Qualitatively, there was light improvement in structures definition in high-resolution image compared to standard resolution image. At the semiquantitative analysis, image reconstruction protocols using high-resolution mode did not significantly improve the recuperation of radiopharmaceutical uptake into analyzed tissues.
Descargas
Referencias
TAN, L. T.; Ong, K. L. Semi-quantitative Measurements of Normal Organs with Variable Metabolic Activity on FDG PET imaging. Ann Acad Med Singapore, v. 33, p. 183–185, 2011.
GONTIJO, R. M. G.; FERREIRA, A. V.; SOUZA, G. C. A.; MENDES, B. M.; SILVA, J. B.; MAMEDE, M. Image quality evaluation of a small animal PET scanner. Brazilian Journal of Radiation Sciences, v. 08-02, p. 01–13, 2020.
BARBOSA J. V. C.; GONTIJO, R. M. G.; SOUZA, G. C. A.; MENDES, B. M.; SILVA, J. B.; MAMEDEB, M.; FERREIRA, A.V. Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C. Brazilian Journal of Radiation Sciences, v. 09-03, p. 01–19, 2019.
NEMA-National ElectricalManufacturers Association. Performance Measurements of Small Animal PET, Rosslyn VA; Standards Publication NU 4-2008, 2008.
OIKONEN, V. Standardized uptake value (SUV), Turku Pet Centre. Turku, Finland. 2020. Available at: <http://www.turkupetcentre.net/petanalysis/model_suv.html>. Last accessed: 18 Jan. 2022.
ADAMS, M. C.; TURKINGTON, T. G.; WILSON, J. M.; WONG, T. Z. A Systematic Review of the Factors Affecting Accuracy of SUV Measurements. American Journal of Roentgenology, v. 195, n. 2, p. 310–320, 2010.
BRENDLE, C.; KUPFERSCHLÄGER, J.; NIKOLAOUA, K.; FOUGÈRE, C.; GATIDIS, S.; PFANNENBERGA, C. Is the standard uptake value (SUV) appropriate for quantification in clinical PET imaging? – Variability induced by different SUV measurements and varying reconstruction methods. European Journal of Radiology. v. 84, n. 1, p. 158-162, 2015.
SAHA, G. B. Basics of PET Imaging: Physics, Chemistry, and Regulations, 2th ed., Springer New York, NY, 2010. p. 83-84.
Descargas
Publicado
Número
Sección
Licencia
Derechos de autor 2023 Brazilian Journal of Radiation Sciences (BJRS)

Esta obra está bajo una licencia internacional Creative Commons Atribución 4.0.
Licencia: los artículos de BJRS tienen una licencia internacional Creative Commons Attribution 4.0, que permite el uso, el intercambio, la adaptación, la distribución y la reproducción en cualquier medio o formato, siempre que se otorgue el crédito correspondiente al autor o autores originales y a la fuente, proporcione un enlace a la licencia Creative Commons e indique si se realizaron cambios. Las imágenes u otros materiales de terceros en el artículo están incluidos en la licencia Creative Commons del artículo, a menos que se indique lo contrario en una línea de crédito al material. Si el material no está incluido en la licencia Creative Commons del artículo y su uso previsto no está permitido por la regulación legal o excede el uso permitido, el autor deberá obtener el permiso directamente del titular de los derechos de autor. Para ver una copia de esta licencia, visite http://creativecommons.org/licenses/by/4.0/