Aspects of an automatic system of implants of radioac-tive seeds and anatomic object simulator for tests in prostate brachytherapy.
DOI:
https://doi.org/10.15392/bjrs.v7i2A.698Palabras clave:
Brachytherapy, implant automatic systems, radioactive seeds.Resumen
This work presents the development of an automatic radioactive seed implantation system (PSIS). PSIS may assist the testing permanent implants procedure in the prostate. These tests will be important in measurements of absorbed doses in the pelvic structures, involving the organs and tissues at risk to improve planning, seed positioning and dosimetry. The automated Prostate Seed Implant System (PSIS) has been designed to meet operational needs, which offers the freedom when positioning the brachytherapy needle inside the treatment area ensuring repeatability and fidelity to the planned treatment. Both the ultrasound probe and the seed implant needle are driven by stepper motors, ATMEGA microcontroller, bearings, aluminum shafts and a GUI (Graphical User Interface). Movement of both the probe and the needle holder was performed by fixed spindle on a threaded rod going through a stepper motors by a coupling. These engines were chosen due to the necessity of movements precision that these types of motors provide. The project was developed and the PSIS prototype was assembled. The prototype presented acceptable operating characteristics for prostate implants. The advantage of this system is that the automation of the application provides an accurate positioning and movement of both probe and seed application. In addition to this study, seeds implantation tests will be performed, and such tests will be essential in protocol validation processes.
Descargas
Referencias
PEREZ C. A., Technical Basis of Radiation Therapy, Springer, p. 275, 2006.
MONTEIRO S. M.; DOMINGAS S. S., IVANILSON S. C., KÉLYA M. M., RAONE S. D. , WANESSA D. B. S., Tratamento Radioterápico do Câncer de Próstata. Revista Científica Multidisciplinar Núcleo do Conhecimento. Ano 2, v. 13. p. 474-481, 2017.
RODRIGUES R. P., Método de aplicação das fontes de iodo-125 radioncologia radiote-rapia, Available at:
<http://ruirodrigues.net/radio2/index.php?option=com_content&view=article&id=135%3Amdo-de-aplica&catid=49%3Abraquiterapia&Itemid=109&showall=1> Last accessed: 25 Jul. 2017.
FRANCA A. C., The seven-year preliminary results of brachytherapy with Iodine-125 seeds for localized prostate cancer treated at a Brazilian single-center. International Braz. J. Urol, v.33, n.6, Rio de Janeiro, 2007.
Driving 28BYJ-48 step motor with Pololu A4988/DRV8825 drivers, Available at:
<http://webcache.googleusercontent.com/search?q=cache:16M90IlcnTIJ:www.electronicsmayhem.com/%3Fp%3D13+&cd=2&hl=pt-BR&ct=clnk&gl=br> Last accessed: 24 Jul. 2017.
Datasheet 28byj48, 12V, Available at:
<http://www.petervis.com/Datasheets/28byj48/28byj48.html> Last accessed: 17 Aug. 2017.
Descargas
Publicado
Número
Sección
Categorías
Licencia
Derechos de autor 2021 Brazilian Journal of Radiation Sciences (BJRS)

Esta obra está bajo una licencia internacional Creative Commons Atribución 4.0.
Licencia: los artículos de BJRS tienen una licencia internacional Creative Commons Attribution 4.0, que permite el uso, el intercambio, la adaptación, la distribución y la reproducción en cualquier medio o formato, siempre que se otorgue el crédito correspondiente al autor o autores originales y a la fuente, proporcione un enlace a la licencia Creative Commons e indique si se realizaron cambios. Las imágenes u otros materiales de terceros en el artículo están incluidos en la licencia Creative Commons del artículo, a menos que se indique lo contrario en una línea de crédito al material. Si el material no está incluido en la licencia Creative Commons del artículo y su uso previsto no está permitido por la regulación legal o excede el uso permitido, el autor deberá obtener el permiso directamente del titular de los derechos de autor. Para ver una copia de esta licencia, visite http://creativecommons.org/licenses/by/4.0/