Monte-Carlo Modelling For Evaluation Of Two Different Calculation Algorithms
DOI:
https://doi.org/10.15392/bjrs.v7i1.792Palabras clave:
3-Dimensional Conformal Radiation Therapy, Anisotropic Analytical Algorithm, Pencil Beam Convolution, Monte CarloResumen
Modeling of linac head (VARIAN Trilogy) for 6 MeV photon beam was performed using BEAMnrc code package (BEAMnrc 2017). The DOSXYZnrc code was used to determine the percentage depth dose (PDD profiles) and beam profiles for different symmetric square field sizes, i.e., 5cm X 5cm; and 40cm X 40cm. The DICOM images of Alderson Radiation Therapy (ART) RANDO Phantom was used. Four field 3D-CRT treatment plans were generated using AAA, PBC, and Monte-Carlo (MC). It was found that nominal energy of 5.7 MeV with FWHM of 1.2 mm provides best matching of modelled and working linac. All three 3D-CRT plans calculated with AAA, PBC and MC on a Pelvic Rando phantom, were compared using CERR (Computational Environment for Radiotherapy Research) and MATLAB 2013b. It was found that AAA and PBC have comparable results, although in case of tissue interfaces and inhomogeneous media AAA provides better accuracy in comparison to PBC. It can also be observed that AAA and PBC underestimate doses in comparison to MC in the soft muscle tissue which includes OARs such as bladder, bowel bag, and PTV TOTAL. It was found that both AAA and PBC fails to account for tissue air interface accurately and shows variation of 30-40% whereas for surface dose variation of +/-10% was observed. In homogeneous media (muscle tissue) AAA and PBC underestimate doses in comparison to MC. These commercially available algorithms overestimates and underestimates dose values as compared to MC based dose calculation for low and high dose regions specially.
Descargas
Referencias
G. M. Ross, “Induction of cell death by radiotherapy”. Endocr. Relat. Cancer, Vol 6, p. 41-44, 1999.
T.C.F. Fonseca, J.P.R. Junior . “MCMEG:Simulations of both PDD and TPR for 6 MV LINAC photon beam using different MC codes”. Radiation Physics and Chemistry. Vol 140; p. 386-391, 2017.
RD Parveen Kumar, KP Santhosh et al., “Estimation of inhomogenity correction factors for Co-60 beam using Monte Carlo Simulations” JCRT Vol. 7 pp. 308, 2011 .
D.W.O. Rogers, B. Walters, I. Kawrakow. BEAMnrc User Manual, National Research Council of Canada, Ottawa & Canada, 2017.
D.W.O. Rogers, B. Walters, et al.. DOSXYZnrc User Manual, National Research Council of Canada, Ottawa & Canada, 2017.
James E. Rodgers. Monte Carlo simulations of dose deposition applied to clinical radiation therapy, Elsevier ,Radiation Measurements 41, 20Q.
I. Kawrakow, Mainegra-Hing, et al. The EGSnrc Code System: Monte Carlo Simulation of Electron and Photon Transport, NRCC Report PIRS-701.
http://www.irs.inms.nrc.ca/BEAM/beamhome.html.
Ryota Onizuka, Fujio Araki, et al. “Accuracy of dose calculation algorithms for virtual heterogeneous phantoms and intensity-modulated radiation therapy in the head and neck”. Radiol Phys Technol, Vol. 9, p. 77, 2016.
Gagné, I. M. and Zavgorodni. “Evaluation of the analytical anisotropic algorithm in an extreme water–lung interface phantom using Monte Carlo dose calculations.” Journal of Applied Clinical Medical Physics Vol. 8, p. 33, 2007).
F Hasenbalg, H Neuenschwander, et al. “Collapsed cone and analytical anisotropic algorithm dose calculations compared to VMC++ Monte Carlo simulations in clinical cases.” J. Phys.: Conf. Ser. 74 p. 21007, 2007.
Descargas
Publicado
Número
Sección
Licencia
Derechos de autor 2021 Brazilian Journal of Radiation Sciences (BJRS)

Esta obra está bajo una licencia internacional Creative Commons Atribución 4.0.
Licencia: los artículos de BJRS tienen una licencia internacional Creative Commons Attribution 4.0, que permite el uso, el intercambio, la adaptación, la distribución y la reproducción en cualquier medio o formato, siempre que se otorgue el crédito correspondiente al autor o autores originales y a la fuente, proporcione un enlace a la licencia Creative Commons e indique si se realizaron cambios. Las imágenes u otros materiales de terceros en el artículo están incluidos en la licencia Creative Commons del artículo, a menos que se indique lo contrario en una línea de crédito al material. Si el material no está incluido en la licencia Creative Commons del artículo y su uso previsto no está permitido por la regulación legal o excede el uso permitido, el autor deberá obtener el permiso directamente del titular de los derechos de autor. Para ver una copia de esta licencia, visite http://creativecommons.org/licenses/by/4.0/