Sum-peak method with two NaI(Tl) crystals: 68(Ge+Ga) standardization
DOI:
https://doi.org/10.15392/bjrs.v7i3B.862Palabras clave:
68Ge, 68Ga, sum-peak method, activity standardization, anticoincidence methodResumen
A 68(Ge+Ga) commercial solution has been standardized in LNMRI in Brazil, by sum-peak method, in which a 3”x3” NaI(Tl) gamma-ray detector is positioned at the top of a well-type 5”x5” NaI(Tl) gamma-ray detector, resulting in a set up approximately 4p geometry. In this work the known germanium volatility was tested using three dried sources and three liquid sources in the sum-peak method measurements and the activity results showed a standard deviation of 0.41%. The activities were compared with another primary method: 4pb-g live-timed anti-coincidence counting. The two methods gave activity concentration values with differences from the certified value of +0.8 % (anticoincidence method) and -3.4% (sum-peak method).
Descargas
Referencias
ZIMMERMAN, B. E.; CESSNA, J. T.; FITZGERALD, R.Standardization of68Ge/68Ga Using Three Liquid Scintillation Counting Based Methods.Journal of Research of the National Insti-tute of Standards and Technology, vol.113, p 265-280, 2008.
SCHÖNFELD,E.DDEP- Decay Data Evaluation Project Data. Available in: www.nucleide.org/DDEP_WG/Nuclides/Ge-68_tables.pdfs. Acessed in: January 1st, 2011.
CHISTÉ, V. and BÉ,M.M., DDEP-Decay Data Evaluation Project Data.Available in: www.nucleide.org/DDEP_WG/Nuclides/F-18_tables.pdfs. Acessed in: January 1st, 2011.
GRIGORESCU, E. L.; NEGUT, C. D.; LUCA, A.; RAZDOLESCU, A. C.; TANASE, M. Standardization of 68(Ge+Ga). Applied Radiation and Isotopes,vol.60, p. 429-431, 2004.
BRINKMAN, G. A.; ATEN JR, A. H. W.; VEENBOER, J. Th. Absolute standardization with a NaI(Tl)crystal—I: calibration by means of a single nuclide. International Journal of Applied Radiation and Isotopes, vol.14, p. 153-157,1963.
BRINKMAN, G. A.; ATEN JR., A. H. W. Absolute Standardization with a NaI(Tl) Crystal-III – Calibration of +-Emitters. International Journal of Applied Radiation and Iso-topes.Vol.14, p. 503-510, 1963.
ALMEIDA, M. C. M.; IWAHARA, A.; POLEDNA, R.; DA SILVA, C. J.; DELGADO J. U. Absolute disintegration rate and 320 keV -ray emission probability of 51Cr. Nuclear Instru-ments and Methods in Physics Research A, vol.580, p. 165-168, 2007.
BRYANT, J. Anticoincidence counting method for standardizing radionuclides emitting de-layed gamma rays. Applied Radiation and Isotopes, vol.13, p. 273-276, 1962.
BAERG, A. P.; MUNZENMAYER, K.; BOWES, G.C. Live-timed anticoincidence counting with extending dead-time circuitry. Metrologia, vol.12, p.77-80, 1976.
DA SILVA, C. J.; IWAHARA, A.; POLEDNA, R., BERNARDES, E. M. O.;DI PRINZIO, M. A. R. R.; LOPES, R. T. Standardization of Ga-67, Cr-51 and Fe-55 by live-timed anti-coincidence counting with extending dead time. Applied Radiation and Isotopesvol. 66, p. 231-235,2008.
DA SILVA, C. J.; IWAHARA, A.; POLEDNA, R.; BERNARDES, E. M. O.; DI PRINZIO, M. A. R. R.; DELGADO, J. U.; LOPES, R. T. Standardization of 241Am, 124Sband 131I by live-timed anticoincidence countingwith extended dead time. Applied Radiation and Isotopes, vol. 66, p. 886-889, 2008.
OLIVEIRA, E. M.; IWAHARA, A.; POLEDNA, R.; DELGADO, J.U.; DA SILVA, C. J.; DA SILVA, R. L.; LOPES, R. T. Standardization of 65Zn by sum-peak method. Applied Radiation and Isotopes, vol. 70, p. 2087 - 2090, 2012.
BOUCHARD, J. A new set of electronic modules (NIM standard) for a coincidence system using the pulse mixing method. Applied Radiation and Isotopes, vol. 56, p. 269-273, 2002.
MAESTRO II, 2002. Software Operator’s Manual. Maestro for Windows.Version 5.33.EG&G. ORTEC.
Descargas
Publicado
Número
Sección
Licencia
Derechos de autor 2019 Brazilian Journal of Radiation Sciences (BJRS)

Esta obra está bajo una licencia internacional Creative Commons Atribución 4.0.
Licencia: los artículos de BJRS tienen una licencia internacional Creative Commons Attribution 4.0, que permite el uso, el intercambio, la adaptación, la distribución y la reproducción en cualquier medio o formato, siempre que se otorgue el crédito correspondiente al autor o autores originales y a la fuente, proporcione un enlace a la licencia Creative Commons e indique si se realizaron cambios. Las imágenes u otros materiales de terceros en el artículo están incluidos en la licencia Creative Commons del artículo, a menos que se indique lo contrario en una línea de crédito al material. Si el material no está incluido en la licencia Creative Commons del artículo y su uso previsto no está permitido por la regulación legal o excede el uso permitido, el autor deberá obtener el permiso directamente del titular de los derechos de autor. Para ver una copia de esta licencia, visite http://creativecommons.org/licenses/by/4.0/