A simple method to back-project isocenter dose of radiotherapy treatments using EPID transit dosimetry
DOI:
https://doi.org/10.15392/bjrs.v5i3.319Palavras-chave:
EPID, portal dosimetry, transit dosimetry, radiotherapyResumo
The aim of this work was to implement a simple algorithm to evaluate isocenter dose in a phantom using the back-projected transmitted dose acquired using an Electronic Portal Imaging Device (EPID) available in a Varian Trilogy accelerator with two nominal 6 and 10 MV photon beams. This algorithm was developed in MATLAB language, to calibrate EPID measured dose in absolute dose, using a deconvolution process, and to incorporate all scattering and attenuation contributions due to photon interactions with phantom. Modeling process was simplified by using empirical curve adjustments to describe the contribution of scattering and attenuation effects. The implemented algorithm and method were validated employing 19 patient treatment plans with 104 clinical irradiation fields projected on the phantom used. Results for EPID absolute dose calibration by deconvolution have showed percent deviations lower than 1%. Final method validation presented average percent deviations between isocenter doses calculated by back-projection and isocenter doses determined with ionization chamber of 1,86% (SD of 1,00%) and -0,94% (SD of 0,61%) for 6 and 10 MV, respectively. Normalized field by field analysis showed deviations smaller than 2% for 89% of all data for 6 MV beams and 94% for 10 MV beams. It was concluded that the proposed algorithm possesses sufficient accuracy to be used for in vivo dosimetry, being sensitive to detect dose delivery errors bigger than 3-4% for conformal and intensity modulated radiation therapy techniques.
Downloads
Referências
Klein EE, Hanley J, Bayouth J, Yin F, Simon W, Dresser S, et al. Task Group 142 report: Quality assurance of medical accelerators. Medical Physics, 2009, 36(9), 4197-4212. doi: 10.1118/1.3190392
World Health Organization. Risk Profile. WHO Press, 2008, Geneve.
International Atomic Energy Agency. IAEA Human Health Reports No. 8: Development of procedures for in vivo dosimetry in radiotherapy. IAEA, 2013, Vienna.
Esch AV, Depuydt T, Huyskens DP. The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields. Radiotherapy and Oncology, 2004, 71(2), 223-234. doi:10.1016/j.radonc.2004.02.018
Greer PB, Popescu CC. Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapy. Medical Physics, 2003, 30(7), 1618-1627. doi:10.1118/1.1582469
Ban L, Chin L, Wronski M, Weiser K, Turner A. Evaluating the Impact of In Vivo EPID Dosimetry on Intensity-Modulated Radiation Therapy Treatment Delivery Workflow: A Stakeholder Perspective. Journal of Medical Imaging and Radiation Sciences, 2014, 45(3), 253-259. doi:10.1016/j.jmir.2013.12.008
Wendling M, Louwe RJ, Mcdermott LN, Sonke J, Herk MV, Mijnheer BJ. Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method. Medical Physics, 2006, 33(2), 259-273. doi:10.1118/1.2147744
Mans A, Wendling M, Mcdermott LN, Sonke J, Tielenburg R, Vijlbrief R, et al. Catching errors with in vivo EPID dosimetry. Medical Physics, 2010, 37(6), 2638-2644. doi:10.1118/1.3397807
Elmpt WV, Mcdermott L, Nijsten S, Wendling M, Lambin P, Mijnheer B. A literature review of electronic portal imaging for radiotherapy dosimetry. Radiotherapy and Oncology, 2008, 88(3), 289-309. doi:10.1016/j.radonc.2008.07.008
Dahlgren CV, Ahnesj A, Montelius A, Rikner G. Portal dose image verification: formalism and application of the collapsed cone superposition method. Physics in Medicine and Biology, 2002, 47(24), 4371-4387. doi:10.1088/0031-9155/47/24/305
Spezi E, Lewis DG. Full forward Monte Carlo calculation of portal dose from MLC collimated treatment beams. Physics in Medicine and Biology, 2004, 49(2), 355-355. doi:10.1088/0031-9155/49/2/c01
Piermattei A, Cilla S, Azario L, Greco F, Russo M, Grusio M, et al. ASi EPIDs for the in-vivo dosimetry of static and dynamic beams. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2015, 796, 93-95. doi:10.1016/j.nima.2015.02.024
Putha S, Saxena P, Banerjee S, Srinivas C, Vadhiraja B, Ravichandran R, et al. On-line estimations of delivered radiation doses in three-dimensional conformal radiotherapy treatments of carcinoma uterine cervix patients in linear accelerator. Journal of Medical Physics, 2016, 41(4), 224. doi:10.4103/0971-6203.195186
Mijnheer B, Olaciregui-Ruiz I, Rozendaal R, Sonke J, Spreeuw H, Tielenburg R, et al. 3D EPID-based in vivo dosimetry for IMRT and VMAT. Journal of Physics: Conference Series, 2013, 444, 012011. doi:10.1088/1742-6596/444/1/012011
Celi S, Costa E, Wessels C, Mazal A, Fourquet A, Francois P. EPID based in vivo dosimetry system: clinical experience and results. Journal of Applied Clinical Medical Physics, 2016, 17(3), 262-276. doi:10.1120/jacmp.v17i3.6070
International Atomic Energy Agency. Technical Report Series No. 398. Absorbed dose determination in external beam radiotherapy. An international code of practice for dosimetry based on standards of absorbed dose to water. IAEA, 2000, Vienna.
Downloads
Publicado
Edição
Seção
Licença
Declaro que o presente artigo é original, não tendo sido submetido à publicação em qualquer outro periódico nacional ou internacional, quer seja em parte ou em sua totalidade. Declaro, ainda, que uma vez publicado na revista Brazilian Journal of Radiation Sciences, editada pela Sociedade Brasileira de Proteção Radiológica, o mesmo jamais será submetido por mim ou por qualquer um dos demais co-autores a qualquer outro periódico. Através deste instrumento, em meu nome e em nome dos demais co-autores, porventura existentes, cedo os direitos autorais do referido artigo à Sociedade Brasileira de Proteção Radiológica, que está autorizada a publicá-lo em meio impresso, digital, ou outro existente, sem retribuição financeira para os autores.
Licença
Os artigos do BJRS são licenciados sob uma Creative Commons Atribuição 4.0 Licença Internacional, que permite o uso, compartilhamento, adaptação, distribuição e reprodução em qualquer meio ou formato, desde que você dê o devido crédito ao (s) autor (es) original (is) e à fonte, forneça um link para a licença Creative Commons, e indique se mudanças foram feitas. As imagens ou outro material de terceiros neste artigo estão incluídos na licença Creative Commons do artigo, a menos que indicado de outra forma em uma linha de crédito para o material. Se o material não estiver incluído no licença Creative Commons do artigo e seu uso pretendido não é permitido por regulamentação legal ou excede o uso permitido, você precisará obter permissão diretamente do detentor dos direitos autorais. Para visualizar uma cópia desta licença, visite http://creativecommons.org/licenses/by/4.0/