Comparison between two methods for estimation of Equilibrium Dose in MSCT procedures
DOI:
https://doi.org/10.15392/bjrs.v10i2A.1793Palavras-chave:
CTDI100, Equilibrium Dose, Computed TomographyResumo
The CTDI100 underestimates the dose indicator due the scattering beyond the scanning length of 100mm of the measurement. A proposed method to obtain the dose (including the “tail” of the dose distribution) is a measurement on a large phantom to estimate the dose indicator on an ideal infinitum large phantom (equilibrium dose). As the scanning length increases, the dose indicator grows asymptotically. Measuremente previously obteined with three tied phantoms of PMMA were adopted in the present work. It includes a set of length measured for different combinations of protocols of voltage, pitch, diameter and axis of measurement. Using an empirically characterized function of the experimental data, two methods to estimate the equilibrium dose were evaluated. One method is manipulating algebraically of the function, and another is the use of Levenberg-Marquardt least-squares. The comparative evaluation of both methods demonstrate that in higher values of the pitch results in a higher difference of both methods. The results demonstrate a good compatibility of the methods for the estimation of the quantity Deq. However, the results for alfa and Leq are significantly different when both methods are considered.
Downloads
Referências
AAPM - American Association of Physicists in Medicine. The Measurement, Reporting, and Management of Radiation Dose in CT. AAPM Report No. 96, Task Group 23. College Park, MD: AAPM, 2008.
DIXON, R. L., The Physics of CT Dosimetry: CTDI and Beyond, 1rd ed., Boca Raton: CRC Press, 2019. p. 5-11.
BOONE, J. M. Trouble with CTDI100, Med Phys, v. 34(4), p. 1364-71, 2007.
AAPM - American Association of Physicists in Medicine. Comprehensive Methodology for the Evaluation of Radiation Dose in X-Ray Computed Tomography. Report No. 111, Task Group 111. College Park, MD: AAPM, 2010.
GONZALES, A. H. L. Simulation and validation of dose profiles and their use to estimate dosimetric quantities for Computed Tomography. 2019. PhD thesis - Physics Institute, Universidade de São Paulo, São Paulo, 2019.
AAPM - American Association of Physicists in Medicine. The Design and Use of the ICRU/AAPM CT Radiation Dosimetry Phantom: An Implementation of AAPM Report 111. AAPM Report No. 200, Task Group 200. Alexandria, VA: AAPM, 2010.
COSTA, P. R., NERSISSIAN, D. Y., UMISEDO, N. K., GONZALES, A. H. L., FERNÁNDEZ- VAREA, J. M. A. comprehensive Monte Carlo study of CT dose metrics proposed by the AAPM Reports 111 and 200. Med Phys, 1–18, 2021.
DIXON, R. L., BALLARD, A. C. Experimental validation of a versatile system of CT dosimetry using a conventional ion chamber: Beyond CTDI 100, Med Phys, v. 34(8), p. 3399-413, 2007.
Downloads
Publicado
Edição
Seção
Categorias
Licença
Direitos autorais (c) 2022 Brazilian Journal of Radiation Sciences

Este trabalho está licenciado sob uma licença Creative Commons Attribution 4.0 International License.
Declaro que o presente artigo é original, não tendo sido submetido à publicação em qualquer outro periódico nacional ou internacional, quer seja em parte ou em sua totalidade. Declaro, ainda, que uma vez publicado na revista Brazilian Journal of Radiation Sciences, editada pela Sociedade Brasileira de Proteção Radiológica, o mesmo jamais será submetido por mim ou por qualquer um dos demais co-autores a qualquer outro periódico. Através deste instrumento, em meu nome e em nome dos demais co-autores, porventura existentes, cedo os direitos autorais do referido artigo à Sociedade Brasileira de Proteção Radiológica, que está autorizada a publicá-lo em meio impresso, digital, ou outro existente, sem retribuição financeira para os autores.
Licença
Os artigos do BJRS são licenciados sob uma Creative Commons Atribuição 4.0 Licença Internacional, que permite o uso, compartilhamento, adaptação, distribuição e reprodução em qualquer meio ou formato, desde que você dê o devido crédito ao (s) autor (es) original (is) e à fonte, forneça um link para a licença Creative Commons, e indique se mudanças foram feitas. As imagens ou outro material de terceiros neste artigo estão incluídos na licença Creative Commons do artigo, a menos que indicado de outra forma em uma linha de crédito para o material. Se o material não estiver incluído no licença Creative Commons do artigo e seu uso pretendido não é permitido por regulamentação legal ou excede o uso permitido, você precisará obter permissão diretamente do detentor dos direitos autorais. Para visualizar uma cópia desta licença, visite http://creativecommons.org/licenses/by/4.0/