Comparison between the response in air and in water of two ionization chambers exposed to low-energy X-rays

Autores

  • Adriana Moreno-Ramírez Instituto de Física, Universidad Nacional Autónoma de México
  • Guerda Massilion ,

DOI:

https://doi.org/10.15392/bjrs.v10i2A.2031

Palavras-chave:

low-energy x-rays, ionization chambers, low-energy dosimetry, measurements in water, measurements in air

Resumo

This work investigated the effect of two different media in the response of two Farmer-type ionization chambers exposed to low-energy x-rays from 27 kV – 155 kV.  The measurements were performed in air and in liquid water. Independent of the ionization chamber, at effective energies below 30 keV (80 kV), the charge collected in liquid water can be up to 84% less than that generated in air. The lower is the energy, the smaller is the response in water compared to air. This can be associated to the rapid attenuation of the low-energy photon fluence by the water depth. At effective energies greater than 30 keV, the response in water becomes larger than that in air, reaching a maximum of 27% and 35% at 65 keV (150 kV) for A19 and A12 chamber, respectively. The difference in response between the two media is consistently greater for ionization chamber A19 at energy above 25 keV.

Downloads

Os dados de download ainda não estão disponíveis.

Referências

MA C.-M., COFFEY C. W., DEWERD L. A., Liu C, NATH R., SELTZER S. M., et al. AAPM protocol for 40–300 kV x-ray beam dosimetry in radiotherapy and radiobiology. Med Phys, v. 28 p. 868-893, 2001.

IAEA. Technical Reports Series No. 398, Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water. Vienna, 2000.

KRAUSS, A.; BÜERMANN, L.; KRAMER, H.-M. and SELBACH, H.-J. Calorimetric determination of the absorbed dose to water for medium-energy x-rays with generating voltages from 70 to 280 kV. Phys Med Biol, v. 57 p. 6245-6268, 2012.

JIMÉNEZ BELTRÁN C. Caracterización del tubo de rayos X YXLON Y.SMART 160E/1.5. B.Sc. Thesis, Instituto Politécnico Nacional, 2015.

Downloads

Publicado

21-07-2022

Como Citar

Comparison between the response in air and in water of two ionization chambers exposed to low-energy X-rays. Brazilian Journal of Radiation Sciences, Rio de Janeiro, Brazil, v. 10, n. 2A (Suppl.), 2022. DOI: 10.15392/bjrs.v10i2A.2031. Disponível em: https://bjrs.org.br/revista/index.php/REVISTA/article/view/2031. Acesso em: 16 jul. 2025.