Evaluation of a Digital Mammography System with the MAM-162 phantom
DOI:
https://doi.org/10.15392/2319-0612.2024.2283Palavras-chave:
DIN 6868-162:2013-06, Quality Control, MAM-162Resumo
Quality controls are performed to verify the performance status of X-ray equipment in relation to image quality and patient dose. Originally, the Laboratory of Radiation Physics and Metrology (LAF-RAM) in Nicaragua had only one phantom to evaluate mammography image quality subjectively; however, under a cooperation project with the International Atomic Energy Agency (IAEA), the MAM-162 phantom was acquired for the evaluation of physical image quality. In this work a CR type digital mammography system was evaluated using the MAM-162 phantom with adaptation to a single quality control protocol based on international standards such as the German Institute for Standardization (DIN) DIN 6868-162:2013-06, the IAEA technical document TECDOC protocol No. 1958 and the European Federation of Organizations for Medical Physics (EFOMP) mammo protocol. Among the main tests performed are the evaluation of the lateral image limit of the chest wall with satisfactory results, the automatic exposure control, the Signal-to-Noise difference ratio that complies with the tolerances, in the case Ghosting and dynamic range the results of both are satisfactory, the presence of artifacts was observed when applying variance mapping, a base value of the dynamic range was also obtained. For the dosimetry, the half value layer and the determination of the Mean Glandular Dose were evaluated, which does not comply with the tolerances. The final protocol document could be validated and implemented as part of the management and service documents of the Quality Control and Monitoring Laboratory (LCM) of LAF-RAM.
Downloads
Referências
Referenzzentrum Mammographie Südwest . Guia de aplicacion para la prueba de aceptación según la normativa en equipos de rayos X para mamografia digital según DIN 6868-162:201306. Alemania. (2013).
COMITE de NORMAS de RADIOLOGIA (NAR). DIN 6868-162; Garantía de calidad de imagen en instalaciones de diagnóstico por rayos X - Parte 162: Prueba de aceptación según la norma en instalaciones de rayos X para mamografía digital. Berlín: Beuth-Verlag, junio de 2013.
OIEA- Organismo Internacional de Energía Atómica. PROTOCOLOS DE CONTROL DE CALIDAD PARA RADIODIAGNÓSTICO EN AMÉRICA LATINA Y EL CARIBE. OIEA- TECDOC-1958. Viena. (2020).
EFOMP-European Federation of Organizations for Medical Physics. EFOMP Mammo Protocol. Quality controls in Digital Mammography. EFOMP. 2017.
ARTINIS. Catálogo MAM-162.
DANCE DR., Monte Carlo calculation of conversion factors for the estimation of mean glandular breast dose Phys. Med. Biol. 35, 1211-1219 (1990)
DANCE DR., SKINNER CL., YOUNG KC., BECKETT JR., KOTRE CJ., Additional factors for the estimation of mean glandular dose using the UK mammography dosimetry protocol, Phys Med. Biol 45:3225-3240 (2000).
DANCE DR., YOUNG KC., VAN ENGEN RE., Further factors for the estimation of mean glandular dose using the United Kingdom, European and IAEA breast dosimetry protocol, Phys Med Biol 54: 4361-4372 (2009).
Downloads
Publicado
Edição
Seção
Categorias
Licença
Direitos autorais (c) 2024 Brazilian Journal of Radiation Sciences

Este trabalho está licenciado sob uma licença Creative Commons Attribution 4.0 International License.
Declaro que o presente artigo é original, não tendo sido submetido à publicação em qualquer outro periódico nacional ou internacional, quer seja em parte ou em sua totalidade. Declaro, ainda, que uma vez publicado na revista Brazilian Journal of Radiation Sciences, editada pela Sociedade Brasileira de Proteção Radiológica, o mesmo jamais será submetido por mim ou por qualquer um dos demais co-autores a qualquer outro periódico. Através deste instrumento, em meu nome e em nome dos demais co-autores, porventura existentes, cedo os direitos autorais do referido artigo à Sociedade Brasileira de Proteção Radiológica, que está autorizada a publicá-lo em meio impresso, digital, ou outro existente, sem retribuição financeira para os autores.
Licença
Os artigos do BJRS são licenciados sob uma Creative Commons Atribuição 4.0 Licença Internacional, que permite o uso, compartilhamento, adaptação, distribuição e reprodução em qualquer meio ou formato, desde que você dê o devido crédito ao (s) autor (es) original (is) e à fonte, forneça um link para a licença Creative Commons, e indique se mudanças foram feitas. As imagens ou outro material de terceiros neste artigo estão incluídos na licença Creative Commons do artigo, a menos que indicado de outra forma em uma linha de crédito para o material. Se o material não estiver incluído no licença Creative Commons do artigo e seu uso pretendido não é permitido por regulamentação legal ou excede o uso permitido, você precisará obter permissão diretamente do detentor dos direitos autorais. Para visualizar uma cópia desta licença, visite http://creativecommons.org/licenses/by/4.0/