Feasibility studies of using thin entrance window photodiodes for clinical electron beam dosimetry
DOI:
https://doi.org/10.15392/bjrs.v7i2A.577Palavras-chave:
Electron dosimetry, Si photodiodes dosimeters, Solid-state dosimetryResumo
The response of the commercial XRA-24 PIN photodiode (5.76 mm2 active area) for clinical electron beam dosimetry covering the range of 8-12 MeV was investigated. Within this energy range, the charge generated in the diode’s sensitive volume is linearly dependent on the absorbed dose up to 320 cGy. However, charge sensitivity coefficients evidenced that the dose response of the diode is slightly dependent on the electron beam energy. Indeed, the diode’s energy dependence was within 8.5% for 8-12MeV electron beams. On the other hand, it was also observed an excellent repeatability of these results with a variation coefficient (VC) lower than 0.4%, which is within the 1% tolerance limit recommended by the AAPM TG-62. Furthermore, the agreement between the percentage depth dose profiles (PDD) gathered with the diode and the ionization chamber allowed achieving the electron beam quality within 1% of that obtained with the ionization chamber. Based on these results, the photodiode XRA-24 can be a reliable and inexpensive alternative for electron beams dosimetry.
Downloads
Referências
GIESSBACH, I.; LAPP, M.; BOHSUNG, J.; GADEMANN, G.; HARDE, D. Dosimetric characteristics of a new unshield silicon diode and its application in clinical photon and electron beams. Medical Physics, v. 32, p. 12, 2005.
ROSENFELD, A. B.; CUTAJAR, D.; LERCH, M. L. F.; TAKACS, G.; CORNELIUS, I. M. ; YUDELEV, M.; ZAIDER, M. Miniature semiconductor detectors for in vivo dosimetry. Radiation Protection Dosimetry, v. 120 (1-4), p. 48-55, 2006.
SONG, H.; AHMAD, M.; DENG, J.; CHEN, Z.; YUE, N. J.; NATH, R. Limitations of silicon diodes for clinical electron dosimetry. Radiation Protection Dosimetry, v. 120 (1-4), p. 56-59, 2006.
SANTOS, T. C.; NEVES, W. F. P.; GONÇALVES, J. A. C.; HADDAD, C. M. K.; HARKONEN, J.; BUENO, C. C. Characterization of miniature RAD-HARD silicon diodes as dosimeters for small fields of photon beams used in radiotherapy. Radiation Measurements, v. 71, p. 396-401, 2014.
GONÇALVES, J. A. C.; PEREIRA, L. N.; POTIENS, M. P. A.; VIVOLO, V.; BUENO, C. C. Evaluation of epitaxial silicon diodes as dosimeters in X-ray mammography. Radiation Measurements, v. 71, p. 384-388, 2014.
BRUZZI, M.; BUCCIOLINI, M.; CASATI, M.; MENICHELLI, D.; TALAMONTI, C.; PIEMEONTE, C.; SVENSSON, B. G. Epitaxial silicon devices for dosimetry applications. Applied Physics Letters, v. 90, p. 172109, 2007.
IAEA. Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water. International Atomic Energy Agency Technical Reports Series (IAEA-TRS) n. 398. Vienna, 2000.
IAEA. The Use of Plane-Parallel Ionization Chambers in High-Energy and Photons Beams: An International Code of Practice for Dosimetry. International Atomic Energy Agency Technical Reports Series (IAEA-TRS) n. 381. Vienna, 1997.
AAPM. Diode in vivo dosimetry for patients receiving external beam radiation therapy. Report of American Association of Physicists in Medicine Radiation Therapy Committee Task Group 62. AAPM- TG-62 Report 87. New York: AAPM, 2005.
SORIANI, A.; DE ANGELIS, C.; ONORI, S.; IACCARINO, G.; MARZI, S.; LANDONI, V.; BENASSI, M. Sistemi a stato sólido per la dosimetria di fasci di elettroni ad alto rateo di dose. In: III Congresso Nazionale dell'Associazione italiana di Fisica in Medicina (AIFM), Agrigento, 2003.
KHOURY, H. J.; HAZIN, C. A.; MASCARENHAS, A. P.; SILVA Jr, E. F. Low cost silicon photodiode for electron dosimetry. Radiation Protection Dosimetry, v. 84 (1-4), p. 341-343, 1999.
Downloads
Publicado
Edição
Seção
Licença
Direitos autorais (c) 2021 Brazilian Journal of Radiation Sciences

Este trabalho está licenciado sob uma licença Creative Commons Attribution 4.0 International License.
Declaro que o presente artigo é original, não tendo sido submetido à publicação em qualquer outro periódico nacional ou internacional, quer seja em parte ou em sua totalidade. Declaro, ainda, que uma vez publicado na revista Brazilian Journal of Radiation Sciences, editada pela Sociedade Brasileira de Proteção Radiológica, o mesmo jamais será submetido por mim ou por qualquer um dos demais co-autores a qualquer outro periódico. Através deste instrumento, em meu nome e em nome dos demais co-autores, porventura existentes, cedo os direitos autorais do referido artigo à Sociedade Brasileira de Proteção Radiológica, que está autorizada a publicá-lo em meio impresso, digital, ou outro existente, sem retribuição financeira para os autores.
Licença
Os artigos do BJRS são licenciados sob uma Creative Commons Atribuição 4.0 Licença Internacional, que permite o uso, compartilhamento, adaptação, distribuição e reprodução em qualquer meio ou formato, desde que você dê o devido crédito ao (s) autor (es) original (is) e à fonte, forneça um link para a licença Creative Commons, e indique se mudanças foram feitas. As imagens ou outro material de terceiros neste artigo estão incluídos na licença Creative Commons do artigo, a menos que indicado de outra forma em uma linha de crédito para o material. Se o material não estiver incluído no licença Creative Commons do artigo e seu uso pretendido não é permitido por regulamentação legal ou excede o uso permitido, você precisará obter permissão diretamente do detentor dos direitos autorais. Para visualizar uma cópia desta licença, visite http://creativecommons.org/licenses/by/4.0/