Verification of angular dependence in MOSFET detector
DOI:
https://doi.org/10.15392/bjrs.v7i2A.734Palavras-chave:
MOSFET, angular dependence, reproducibility.Resumo
In vivo dosimetry is an essential tool for quality assurance programs, being a procedure commonly performed with thermoluminescent dosimeters (TLDs) or diodes. However, a type of dosimeter that has increasing popularity in recent years is the metal-oxide-semiconductor field effect transistor (MOSFET) detector. MOSFET dosimeters fulfill all the necessary characteristics to realize in vivo dosimetry since it has a small size, good precision and feasibility of meas-urement, as well as easy handling. Nevertheless, its true differential is to allow reading of dose in real time, enabling immediate intervention in the correction of physical parameters deviations and anticipation of small anatomical changes in the patient during treatment. In order for MOSFET dosimeter to be better accepted in clinical routine, information reporting performance should be available frequently. For this reason, this work proposes to verify reproducibility and angular dependence of a standard sensitivity MOSFET dosimeter (TN-502RD-H) for Cs-137 and Co-60 sources. Exper-imental data were satisfactory and MOSFET dosimeter presented a reproducibility of 3.3% and 2.7% (±1 SD) for Cs-137 and Co-60 sources, respectively. In addition, an angular dependence of up to 6.1% and 16.3% for both radioactive sources, respectively. It is conclusive that MOSFET dosimeter TN-502RD-H has satisfactory reproducibility and a considerable angular dependence, mainly for the Co-60 source. This means that although precise measurements, special attention must be taken for applications in certain anatomical regions in a patient.Downloads
Referências
WHO – World Health Organization. World Health Organization: Who is cancer? Availa-ble at: <http://www.who.int/cancer/en/>. Last accessed: 01 Dec. 2017.
INCA – Instituto Nacional do Câncer. Instituto Nacional do Câncer: Tratamento do Câncer. Available at <http://www2.inca.gov.br/wps/wcm/connect/cancer/site/tratamento> Last ac-cessed: 01 Dec. 2017.
INCA – Instituto Nacional do Câncer. Instituto Nacional do Câncer: Radioterapia. Avai-lable at <http://www.inca.gov.br/conteudo_view.asp?ID=100> Last accessed: 01 Sept. 2017.
SCALCHI, P., FRANCESCON, P., Calibration of a MOSFET detection system for 6-MV in-vivo dosimetry. Int. J. Radiation Oncology Biol. Phys., v. 40, p. 987-993, 1998.
GOPIRAJ, A.; BILLIMAGGA, R. S.; RAMASUBRAMANIAN, V. Performance characte-ristics and commissioning of MOSFET as an in-vivo dosimeter for high energy photon external be-am radiation therapy. Reports of Practical Oncology & Radiotherapy, v. 13, p. 114-125, 2008.
BOWER, M. W. Physical anthropomorphic phantom of a one-year-old child with real-time dosimetry, Gainesville, Florida, United States of America, 1997.
RAMANI, R.; RUSSEL, S.; O’BRIEN, P. Clinical dosimetry using MOSFETs, Int. J. Ra-diation Oncology Biol. Phys., v 37, p.959-964, 1997.
RAMASESHAN, R.; KOHLI, K. S.; ZHANG, T. J.; LAM, T.; NORLINGER, B.; HAL-LIL, A.; ISLAM, M. Performance characteristics of a microMOSFET as an in vivo dosimeter in radiation therapy, Phys. Med. Biol., v 49, p. 4031-4048, 2004.
SHARP, R.E.; PATER, S.L. The use of pMOS dosimeters at megagray total doses, AEA Technol, RADECS, 1995.
SOUBRA, M.; CYGLER, J.; MACKAY, G. Evaluation of dual bias dual metal oxide-silicon semiconductor field effect transistor detector as radiation dosimeter, Med. Phys., v 21, p. 567-572, 1994.
BEST MEDICAL CANADA. Best Medical Canada: MOSFET Dosimeters. Available at: <http://www.bestmedicalcanada.com/product_dosimeters.html> Last accessed: 01 Dec. 2017.
SILVA JR, I.A. Desenvolvimento e Implementação de um Sistema Automatizado para Adequação do Processo de Calibração de Monitores de Radiação Gama, Universidade de São Paulo, São Paulo, Brazil, 2012.
KUMAR, A. S. et al. Characteristics of mobile MOSFET dosimetry system for megavoltage photon beams. J. Med Phys. v. 39, p. 142-149, 2014.
BEST MEDICAL CANADA. Operator’s manual for Mobile MOSFET System. Ottawa, Canada, 2008.
SCALCHI, P. et al. Characterization of a new MOSFET detector configuration for in vivo skin dosimetry. Medical Physics, v. 32, p. 1571-1578, 2005.
Downloads
Publicado
Edição
Seção
Categorias
Licença
Direitos autorais (c) 2021 Brazilian Journal of Radiation Sciences

Este trabalho está licenciado sob uma licença Creative Commons Attribution 4.0 International License.
Declaro que o presente artigo é original, não tendo sido submetido à publicação em qualquer outro periódico nacional ou internacional, quer seja em parte ou em sua totalidade. Declaro, ainda, que uma vez publicado na revista Brazilian Journal of Radiation Sciences, editada pela Sociedade Brasileira de Proteção Radiológica, o mesmo jamais será submetido por mim ou por qualquer um dos demais co-autores a qualquer outro periódico. Através deste instrumento, em meu nome e em nome dos demais co-autores, porventura existentes, cedo os direitos autorais do referido artigo à Sociedade Brasileira de Proteção Radiológica, que está autorizada a publicá-lo em meio impresso, digital, ou outro existente, sem retribuição financeira para os autores.
Licença
Os artigos do BJRS são licenciados sob uma Creative Commons Atribuição 4.0 Licença Internacional, que permite o uso, compartilhamento, adaptação, distribuição e reprodução em qualquer meio ou formato, desde que você dê o devido crédito ao (s) autor (es) original (is) e à fonte, forneça um link para a licença Creative Commons, e indique se mudanças foram feitas. As imagens ou outro material de terceiros neste artigo estão incluídos na licença Creative Commons do artigo, a menos que indicado de outra forma em uma linha de crédito para o material. Se o material não estiver incluído no licença Creative Commons do artigo e seu uso pretendido não é permitido por regulamentação legal ou excede o uso permitido, você precisará obter permissão diretamente do detentor dos direitos autorais. Para visualizar uma cópia desta licença, visite http://creativecommons.org/licenses/by/4.0/