Calibration and evaluation of the MTS-N dosimeter for absorbed dose measurement in blood components
DOI:
https://doi.org/10.15392/2319-0612.2024.2502Palabras clave:
Blood irradiation, Dosimetry, Calibration, ThermoluminescenceResumen
The use of ionizing radiation has significantly expanded, particularly in the healthcare sector, where the irradiation of blood components prevents transfusion-associated graft-versus-host disease (TA-GvHD). The ISO ASTM 51939:2017 standard establishes a minimum dose of 25 Gy for this procedure, requiring a precise and traceable dosimetric system. This study investigates the calibration of MTS-N thermoluminescent dosimeters at the Gamma Irradiation Laboratory (LIG) of CDTN, which currently uses radiochromic films for routine dosimetry. The dosimeters were calibrated using a RISØ TL/OSL DA-20 reader, and the results indicated that doses above 5 Gy caused reader saturation, distorting the TL emission curve. The introduction of attenuating filters maintained linearity up to 25 Gy, and the calibration coefficient was determined as 6.774×10⁻⁷ Gy/Count, confirming the system’s suitability for dosimetry in blood irradiation at LIG/CDTN.
Descargas
Referencias
[1] Pino, E. S., Giovedi, C. "Ionizing Radiation and Its Applications in Industry," UNILUS Teaching and Research Journal, no. 2, vol. 2, Jan/Jun 2015.
[2] American Association Blood Bank. Irradiation. Available at: http://www.aabb.org/advocacy/regulatorygovernment/irradiation/Pages/default.aspx. Accessed on: 27 Feb. 2024.
[3] ALAN, H. Y. et al. A study on quality control check of dose distribution in blood bags irradiated by 137Cs source blood irradiator by using EBT3 radiochromic film dosimetry. Radiation Physics and Chemistry, v. 214, p. 111303, 2024.
[4] ISO/ASTM. (2017). ISO/ASTM 51939:2017 - Standard Practice for Blood Irradiation Dosimetry.
[5] CALHEIRO, D. S. et al. Mapping absorbed dose distribution in containers of blood and blood components using radiochromic dosimeters irradiated in an industrial irradiator. Radiation Physics and Chemistry, v. 222, p. 111797, 2024.
[6] GRASSO, S. et al. Absorbed dose evaluation of a blood irradiator with alanine, TLD-100 and ionization chamber. Applied Radiation and Isotopes, v. 200, p. 110981, 2023.
Descargas
Publicado
Número
Sección
Categorías
Licencia
Derechos de autor 2024 Lorena Fonseca, Daniel Silva Calheiro, Amir Mesquita, Thêssa Alonso, Peterson Squair

Esta obra está bajo una licencia internacional Creative Commons Atribución 4.0.
Licencia: los artículos de BJRS tienen una licencia internacional Creative Commons Attribution 4.0, que permite el uso, el intercambio, la adaptación, la distribución y la reproducción en cualquier medio o formato, siempre que se otorgue el crédito correspondiente al autor o autores originales y a la fuente, proporcione un enlace a la licencia Creative Commons e indique si se realizaron cambios. Las imágenes u otros materiales de terceros en el artículo están incluidos en la licencia Creative Commons del artículo, a menos que se indique lo contrario en una línea de crédito al material. Si el material no está incluido en la licencia Creative Commons del artículo y su uso previsto no está permitido por la regulación legal o excede el uso permitido, el autor deberá obtener el permiso directamente del titular de los derechos de autor. Para ver una copia de esta licencia, visite http://creativecommons.org/licenses/by/4.0/