Use of an anthropomorphic hand phantom to verify the radiation intensity that is needed to modify the analog and digital radiographic quality

Authors

  • Caroline Kretezel Bandeira Universidade Tecnológica Federal do Paraná Instituto Federal do Paraná
  • Michele Patrícia Müller Mansur Vieira Instituto Federal do Paraná
  • José Eduardo dos Reis Felix Universidade Federal do Triângulo Mineiro

DOI:

https://doi.org/10.15392/bjrs.v3i1A.169

Keywords:

phantoms, proteção radiológica, controle de qualidade

Abstract

The radiology is a field of medicine that is in constantly expansion and advancing. This can be noticed with the transition from analog to digital radiology systems, it is important that professionals understand image formation in both systems in order to produce radiographies with diagnostic quality. Therefore, the objective of this work is to present the importance of radiological protection by changing values of technical parameters while the quality of radiographic imaging is sustained. An anthropomorphic hand phantom was built in order to obtain radiographies as it is necessary to respect the Brazilian regulations (Portaria 453/98) which forbids the use of radiation in patients for testing. Three analog and eight digital radiographies were obtained using fixed kVp and varying mAs. Each image was compared to the others acquired in the same location. Digital radiographies have shown that approximately 28% of change in mAs is necessary to change noise, whereas approximately 33,3% is necessary in the analog system to change density. The conclusion is that computerized systems need less x-ray intensity to modify image features and moreover can reduce the patient radiation doses. However, more testing must be conducted in different radiologic environments to confirm the results obtained in the present study. 

Downloads

Download data is not yet available.

References

BONTRAGER, K. L.; LAMPIGNANO, J. P. Tratado de posicionamento radiográfico e ana-tomia associada.7th ed., Rio de Janeiro: Elsevier, 2010.

BRASIL. Ministério da Saúde. Secretaria de Vigilância Sanitária. Portaria nº 453, de 01º de ju-nho de 1998. Aprova o Regulamento Técnico que estabelece as diretrizes básicas de proteção radiológica em radiodiagnóstico médico e odontológico, dispõe sobre o uso dos raios-x diagnós-ticos em todo território nacional e dá outras providências. Diário Oficial da União da República Federativa do Brasil, Brasília, 02 de junho de 1998.

BUSHONG, S. C. Ciência radiológica para tecnólogos, 9th ed., Rio de Janeiro: Elsevier, 2010.

DO VAL, F. L. Manual de técnica radiográfica. São Paulo: Manole, 2006.

ICRU - International Commission on Radiation Units and Measurements. Tissue substitutes in Radiation Dosimetry and Measurement. ICRU Report 44, Bethesda: ICRU, 1989. 22p.

Published

2015-05-23

How to Cite

Bandeira, C. K., Müller Mansur Vieira, M. P., & dos Reis Felix, J. E. (2015). Use of an anthropomorphic hand phantom to verify the radiation intensity that is needed to modify the analog and digital radiographic quality. Brazilian Journal of Radiation Sciences, 3(1A (Suppl.). https://doi.org/10.15392/bjrs.v3i1A.169