Evaluation of a Digital Mammography System with the MAM-162 phantom
DOI:
https://doi.org/10.15392/2319-0612.2024.2283Keywords:
DIN 6868-162:2013-06, Quality Control, MAM-162Abstract
Quality controls are performed to verify the performance status of X-ray equipment in relation to image quality and patient dose. Originally, the Laboratory of Radiation Physics and Metrology (LAF-RAM) in Nicaragua had only one phantom to evaluate mammography image quality subjectively; however, under a cooperation project with the International Atomic Energy Agency (IAEA), the MAM-162 phantom was acquired for the evaluation of physical image quality. In this work a CR type digital mammography system was evaluated using the MAM-162 phantom with adaptation to a single quality control protocol based on international standards such as the German Institute for Standardization (DIN) DIN 6868-162:2013-06, the IAEA technical document TECDOC protocol No. 1958 and the European Federation of Organizations for Medical Physics (EFOMP) mammo protocol. Among the main tests performed are the evaluation of the lateral image limit of the chest wall with satisfactory results, the automatic exposure control, the Signal-to-Noise difference ratio that complies with the tolerances, in the case Ghosting and dynamic range the results of both are satisfactory, the presence of artifacts was observed when applying variance mapping, a base value of the dynamic range was also obtained. For the dosimetry, the half value layer and the determination of the Mean Glandular Dose were evaluated, which does not comply with the tolerances. The final protocol document could be validated and implemented as part of the management and service documents of the Quality Control and Monitoring Laboratory (LCM) of LAF-RAM.
- Views: 66
- PDF Downloads: 63
- XML Downloads: 4
Downloads
References
Referenzzentrum Mammographie Südwest . Guia de aplicacion para la prueba de aceptación según la normativa en equipos de rayos X para mamografia digital según DIN 6868-162:201306. Alemania. (2013).
COMITE de NORMAS de RADIOLOGIA (NAR). DIN 6868-162; Garantía de calidad de imagen en instalaciones de diagnóstico por rayos X - Parte 162: Prueba de aceptación según la norma en instalaciones de rayos X para mamografía digital. Berlín: Beuth-Verlag, junio de 2013.
OIEA- Organismo Internacional de Energía Atómica. PROTOCOLOS DE CONTROL DE CALIDAD PARA RADIODIAGNÓSTICO EN AMÉRICA LATINA Y EL CARIBE. OIEA- TECDOC-1958. Viena. (2020).
EFOMP-European Federation of Organizations for Medical Physics. EFOMP Mammo Protocol. Quality controls in Digital Mammography. EFOMP. 2017.
ARTINIS. Catálogo MAM-162.
DANCE DR., Monte Carlo calculation of conversion factors for the estimation of mean glandular breast dose Phys. Med. Biol. 35, 1211-1219 (1990) DOI: https://doi.org/10.1088/0031-9155/35/9/002
DANCE DR., SKINNER CL., YOUNG KC., BECKETT JR., KOTRE CJ., Additional factors for the estimation of mean glandular dose using the UK mammography dosimetry protocol, Phys Med. Biol 45:3225-3240 (2000). DOI: https://doi.org/10.1088/0031-9155/45/11/308
DANCE DR., YOUNG KC., VAN ENGEN RE., Further factors for the estimation of mean glandular dose using the United Kingdom, European and IAEA breast dosimetry protocol, Phys Med Biol 54: 4361-4372 (2009). DOI: https://doi.org/10.1088/0031-9155/54/14/002
Published
How to Cite
Issue
Section
Categories
License
Copyright (c) 2024 Brazilian Journal of Radiation Sciences
This work is licensed under a Creative Commons Attribution 4.0 International License.
Licensing: The BJRS articles are licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/