Reproduction of a left-breast 3DCRT field-in-field radiation therapy planning in an anthropomorphic and anthropometric phantom
DOI:
https://doi.org/10.15392/bjrs.v7i2A.659Keywords:
breast teletherapy, breast phantom, dosimetry.Abstract
The proposal of this study was to reproduce the dosimetry established by a treatment planning system (TPS) following a 3D conformational radiation therapy (3DCRT) protocol of two parallel-opposite fields applied to the left-breast in a thorax phantom, with the complementary use of the field-in-field technique. The computed tomography (CT) images of the thorax phantom were transferred to the XiO version-5 for the elaboration of the breast teletherapeutic planning with 2 Gy per day, in 25 fractions, with prescribed dose of 50 Gy. A set of ten EBT2 radiochromic films were irradiated at doses gradually. The values of RGB (Red, Green, Blue) of the films were obtained by scanning and data transformed in optical density (OD), whose values were used to construct the calibration curve. Radiochromic films were positioned outside and inside of the thorax phantom. After phantom radiation at the linear accelerator 6 MV Elekta Precise were measured the intensities of the films in RGB by software ImageJ, transformed in OD and converted in bidimensional dose distributions, applying the calibration curve. In addition, graphics and dose-volume histograms (DVH) were developed. The dose measurements in the glandular-TE in breast did not present statistically significant differences in relation to values at equivalent positions generated in the TPS. The organs at risk received doses below the reference values, according to TPS. It was possible to reproduce the dosimetry prescribed in TPS into the thorax phantom exposed to a breast conformational teletherapy. The use of the radiochromic films in dosimetry proves to be increasingly useful and shall become routine in radiotherapy services.
- Views: 102
- PDF Downloads: 109
Downloads
References
INCA- Instituto Nacional de Câncer. Estimativa 2016-2017: Incidência de Câncer no Brasil. Rio de Janeiro, 2015.
THOMPSON, L.; DIAS, H. G.; CAMPOS, T. P. R. Dosimetry in brain tumor phantom at 15 MV 3D conformal radiation therapy. Radiation Oncology (Online), v. 8, p.168, 2013.
XU, X. G. Computational phantoms for radiation dosimetry: a 40-Year history of evo-lution. In Handbook of Anatomical Models for Radiation Dosimetry, Edited by Xu XG, Ecker-man KF, Boca Raton: Taylor & Francis, 2013.
HOMMA, L. A. H. Aspects of the application of sentinel lymph node surgery and com-plementary brachytherapy in initial invasive breast cancer, Master's Dissertation, Nuclear En-gineering Department, Federal University of Minas Gerais, Brazil, 2005.
NOGUEIRA, L. B.; SILVA, H. L. L.; CAMPOS, T. P. R. Experimental dosimetry in con-formal breast teletherapy compared with the planning system. Applied Radiation and Isotopes, v. 97, p. 93-100, 2015.
NOGUEIRA, L. B.; SILVA, H. L. L.; SILVA, S. D.; CAMPOS, T. P. R. Skin dosimetry in breast teletherapy on an anthropomorphic and anthropometric phantom. Austin J Radiol., v. 2(4), p.1024, 2015.
VOLKER, R..; ABDUL, A.; AZIZ, A. L.; et al. Tangential beam IMRT versus tangential beam 3D-CRT of the chest wall in postmastectomy breast cancer patients: a dosimetric comparison. Radiation Oncology, v. 6, p. 392-406, 2011.
SALVAJOLI, J. V.; SOUHANI, L.; FARIA, S. L. The Role of Radiotherapy in Cancer Treatment - Advances and Challenges, Radioterapia em oncologia, 2nd ed, Rio de Janeiro: Medsi Editora Medica e Científica, 2013.
Ministry of Health- Ordinance-453 Guidelines for Radiological Protection in Medical and Dental Radiology. Secretariat of Sanitary Monitoring, Brazil, 1998.
BUTSON, M. J.; CHEUNG, T.; YU, P. K. Radiochromic film dosimetry in water phantoms. Phys Med Biol., v. 46(1), p. 24-31, 2001.
AMARAL, L. L.; OLIVEIRA, H. F.; FAIRBANKS, L. R.; NICOLUCCI, NETTO, T.G. Averification methodology for in vivo dosimetry in stereotactic radiotherapy. Braz.J.Med. Phys, v.6(3), p. 129–132, 2012.
CAMPOS, T. P. R; THOMPSON, L.; NOGUEIRA, L. B.; DUARTE, I. L.; MATOS, A. S.; TEIXEIRA, C. H; et. al. Anthropomorphic and Anthropometric Simulators of the Structures, Tissues and Organs of the Human Body, Brazil patent BR PI1004465-5, UFMG, Minas Gerais, Brazil, 2012.
SCHETTINI, M. P.; MAIA, M.; CAMPOS, T. P. R. The Development of an anthropo-morphic and anthropometric thorax female phantom for experimental radiodosimentry. Int J Low Radiat., v. 4, p. 124-135, 2007.
ICRU 44- International Commission on Radiation Units and Measurements. Tissue substi-tutes in radiation dosimetry and measurement. ICRU Report 44, Bethesda: ICRU, 1989.
ISP. International Specialty Products. Gatchromic self- developing dosimetry films | EBT2 | EBT 3 | Cyberknife | HD-V2 | MD-V3 | RTQA2 Ashland Inc. New Jersey: ISP, 2010.
IAEA- International Atomic Energy Agency. Absorbed dose determination in external beam radiotherapy: An Internationa code of practice for dosimetry based on standards of ab-sorbed dose to water. Vienna: IAEA , 2000.
RASBAND, W. S. ImageJ, U. S. National Institutes of Health. Bethesda, Maryland, USA. 2011. Available at: <http://imagej.nih.gov/ij/>. Last accessed: jun. 2017.
NOGUEIRA, L. B.; CAMPOS, T. P. R. Nuclear characterization and investigation of radio-active bioglass seed surfaces for brachytherapy via scanning electron microscopy. Journal of Sol-Gel Science and Technology, v. 58, p. 251 – 258, 2011.
ICRU 62- International Commission on Radiation Units. Prescribing, Recording and Re-porting Photon Beam Therapy. ICRU Report 62, Bethesda: ICRU, 1999.
ICRU 50- International Commission on Radiation. Prescribing, Recording, and Reporting Photon Beam Therapy. ICRU Report 50, Bethesda: ICRU, 1993.
MARKS, L. B.; YORKE, E. D.; JACKSON, A.; TENHAKEN, R. K.; CONSTINE, L. S.; EISBRUCH, A. Use of normal tissue complication probability models in the clinic. Int.J. Radiat. Oncol. Biol.Phys, v. 76(3), p. 10–19, 2010.
ORIGINLAB: Origin and OriginPro - Data analysis and graphing – Origin. Available at: <http://www.originlab.com/> . Last accessed: jun. 2017.
Published
How to Cite
Issue
Section
Categories
License
Copyright (c) 2021 Brazilian Journal of Radiation Sciences
This work is licensed under a Creative Commons Attribution 4.0 International License.
Licensing: The BJRS articles are licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/