Verification of angular dependence in MOSFET detector

Authors

  • Clayton Henrique de Souza Instituto de Pesquisas Energéticas e Nucleares - IPEN / CNEN

DOI:

https://doi.org/10.15392/bjrs.v7i2A.734

Keywords:

MOSFET, angular dependence, reproducibility.

Abstract

In vivo dosimetry is an essential tool for quality assurance programs, being a procedure commonly performed with thermoluminescent dosimeters (TLDs) or diodes. However, a type of dosimeter that has increasing popularity in recent years is the metal-oxide-semiconductor field effect transistor (MOSFET) detector. MOSFET dosimeters fulfill all the necessary characteristics to realize in vivo dosimetry since it has a small size, good precision and feasibility of meas-urement, as well as easy handling. Nevertheless, its true differential is to allow reading of dose in real time, enabling immediate intervention in the correction of physical parameters deviations and anticipation of small anatomical changes in the patient during treatment. In order for MOSFET dosimeter to be better accepted in clinical routine, information reporting performance should be available frequently. For this reason, this work proposes to verify reproducibility and angular dependence of a standard sensitivity MOSFET dosimeter (TN-502RD-H) for Cs-137 and Co-60 sources. Exper-imental data were satisfactory and MOSFET dosimeter presented a reproducibility of 3.3% and 2.7% (±1 SD) for Cs-137 and Co-60 sources, respectively. In addition, an angular dependence of up to 6.1% and 16.3% for both radioactive sources, respectively. It is conclusive that MOSFET dosimeter TN-502RD-H has satisfactory reproducibility and a considerable angular dependence, mainly for the Co-60 source. This means that although precise measurements, special attention must be taken for applications in certain anatomical regions in a patient.

Downloads

Download data is not yet available.

Author Biography

Clayton Henrique de Souza, Instituto de Pesquisas Energéticas e Nucleares - IPEN / CNEN

Centro de Engenharia Nuclear

References

WHO – World Health Organization. World Health Organization: Who is cancer? Availa-ble at: <http://www.who.int/cancer/en/>. Last accessed: 01 Dec. 2017.

INCA – Instituto Nacional do Câncer. Instituto Nacional do Câncer: Tratamento do Câncer. Available at <http://www2.inca.gov.br/wps/wcm/connect/cancer/site/tratamento> Last ac-cessed: 01 Dec. 2017.

INCA – Instituto Nacional do Câncer. Instituto Nacional do Câncer: Radioterapia. Avai-lable at <http://www.inca.gov.br/conteudo_view.asp?ID=100> Last accessed: 01 Sept. 2017.

SCALCHI, P., FRANCESCON, P., Calibration of a MOSFET detection system for 6-MV in-vivo dosimetry. Int. J. Radiation Oncology Biol. Phys., v. 40, p. 987-993, 1998.

GOPIRAJ, A.; BILLIMAGGA, R. S.; RAMASUBRAMANIAN, V. Performance characte-ristics and commissioning of MOSFET as an in-vivo dosimeter for high energy photon external be-am radiation therapy. Reports of Practical Oncology & Radiotherapy, v. 13, p. 114-125, 2008.

BOWER, M. W. Physical anthropomorphic phantom of a one-year-old child with real-time dosimetry, Gainesville, Florida, United States of America, 1997.

RAMANI, R.; RUSSEL, S.; O’BRIEN, P. Clinical dosimetry using MOSFETs, Int. J. Ra-diation Oncology Biol. Phys., v 37, p.959-964, 1997.

RAMASESHAN, R.; KOHLI, K. S.; ZHANG, T. J.; LAM, T.; NORLINGER, B.; HAL-LIL, A.; ISLAM, M. Performance characteristics of a microMOSFET as an in vivo dosimeter in radiation therapy, Phys. Med. Biol., v 49, p. 4031-4048, 2004.

SHARP, R.E.; PATER, S.L. The use of pMOS dosimeters at megagray total doses, AEA Technol, RADECS, 1995.

SOUBRA, M.; CYGLER, J.; MACKAY, G. Evaluation of dual bias dual metal oxide-silicon semiconductor field effect transistor detector as radiation dosimeter, Med. Phys., v 21, p. 567-572, 1994.

BEST MEDICAL CANADA. Best Medical Canada: MOSFET Dosimeters. Available at: <http://www.bestmedicalcanada.com/product_dosimeters.html> Last accessed: 01 Dec. 2017.

SILVA JR, I.A. Desenvolvimento e Implementação de um Sistema Automatizado para Adequação do Processo de Calibração de Monitores de Radiação Gama, Universidade de São Paulo, São Paulo, Brazil, 2012.

KUMAR, A. S. et al. Characteristics of mobile MOSFET dosimetry system for megavoltage photon beams. J. Med Phys. v. 39, p. 142-149, 2014.

BEST MEDICAL CANADA. Operator’s manual for Mobile MOSFET System. Ottawa, Canada, 2008.

SCALCHI, P. et al. Characterization of a new MOSFET detector configuration for in vivo skin dosimetry. Medical Physics, v. 32, p. 1571-1578, 2005.

Published

2019-02-25

How to Cite

Souza, C. H. de. (2019). Verification of angular dependence in MOSFET detector. Brazilian Journal of Radiation Sciences, 7(2A (Suppl.). https://doi.org/10.15392/bjrs.v7i2A.734