Response verification of integrated device used for radon measurements in air

Authors

  • Mounir AÏT-ZIANE Laboratoire de Dosimétrie, Centre de Recherche Nucléaire d’Alger, BP. 399 Alger RP, Algiers
  • Malika ALLAB Laboratoire SNIRM, Université des Sciences et de la Technologie Houari Boumediene, BP. 32 Bab Ezzouar, Algiers
  • Zohra LOUNS-MOKRANI Laboratoire de Dosimétrie, Centre de Recherche Nucléaire d’Alger, BP. 399 Alger RP, Algiers

DOI:

https://doi.org/10.15392/bjrs.v7i3.859

Keywords:

radon, electrets, radon monitor

Abstract

In order to establish the national radon level according to the ICRP recommendations, radon levels should be estimated in different living environments (dwellings, workplaces, hospitals, etc.). A particular attention should be given to the correctness of the radon concentration measurement devices in order to maintain their traceability to primary standard laboratory. Radon monitor systems have thus been calibrated using a NIST standard source which is considered as an important step in standardizing radon measurement methods.

In this paper, we describe the system used in our laboratory for the verification against radon response of different radon monitors such as Electret Ion Chamber (EIC) and AlphaGUARD continuous radon monitor (AG), and then we present comparison between these two measurements devices used in some locations at the Centre de Recherche Nucléaire d’Alger (CRNA). As results, EIC system produces results comparable to those produces by the AG system. 

Downloads

Download data is not yet available.

References

SWEDJEMARK, G.A.; MJÖNES, L. Radon and radon daughter concentrations in Swedish homes. Radiation Protection Dosimetry, v. 7, p. 341–345, 1983

UNSCEAR- United Nations Scientific Committee on the Effects of Atomic Radiation. Sources and effects of ionizing radiation, Report to the General Assembly, United Nations, New York, UNSCEAR 2000

MILES, J.C.H.; APPLETON, J.D.; REES, D.M.; GREEN, B.M.R.; ADLAM, K.A.M.; MYERS, A.H. Indicative atlas of radon in England and Wales, Health Protection Agency and British Geological Survey, 2007

VILLALBA, L.; MONTERO-CABRERA, M.E.; MANJON-COLLADO, G.; COLMENERO-SUJO, L.; RENTERIA-VILLALOBOS, M.; CANO- JIMENEZ, A.; RODRIGUEZ-PINEDA, A.; DAVILA-RANGEL, I.; QUIRINO-TORRES, L.; HERRERA-PERAZA, E.F. Natural radioactivity in groundwater and estimates of committed effective dose due to water ingestion in the state of Chihuahua (Mexico). Radiation Protection Dosimetry, v. 121(2), p. 148–157, 2006

JANIK, M.; TOKONAMI, S.; KOVÁCS, T.; KÁVÁSI, N.; KRANROD, C.; SORIMACHI, A. International intercomparisons of integrating radon detectors in the NIRS radon chamber, Applied Radiation and Isotopes, v. 67, p. 1691–1696, 2009

ENSDF (2005) Evaluated Nuclear Structure Data File Search and Retrieval, https://www.nndc.bnl.gov/ensdf/

KOTRAPPA, P.; DEMPSEY, J.C.; HICKEY, J.R; STIEFF, L.R. An electret passive environmental 222Rn monitor based on ionization measurement. Health Physics, v. 54(1), p. 47–56, 1988

KOTRAPPA P.; DEMPSEY J.C.; RAMSEY R.W.; STIEFF L.R. A practical E-Perm system for indoor 222Rn measurement, Health Physics, v. 58, p. 461–466, 1990

KOTRAPPA, P. E-PERM electret ion chambers for measuring radon in water, E-PERM System Manual, Rad Elec Inc. 1–13, 1999

BÜRKIN, W. Calibration using the NIST standard reference material SRM 4973 (Brief Tutorial), GENITRON System Manual, Saphymo GmbH, 2009

NIST-National Institute of Standards and Technology, Certificate for Standard Reference Material SRM-4973, 222Rn Emanation standard, NIST 2005

AL MANSY, M.K.; EL-SAMMAN, H.; HASSAN, M.A.; MAHFOUZ, E.; EISSA, H.; and EL-FIKI, M. Effect of variation in the ambient temperature on the NIST encapsulated 222Rn emanation standard source, In: Proceeding of The International Conference for Environmental Hazard Mitigation (ICEHM 2000), Cairo, Egypt, 9–12 September 2000, 576- 579

EPA-Environmental Protection Agency, The national measurement proficiency (RMP) program, Handbook. EPA520/1-90-006 Feb 1991, First edition, EPA 1991

DENMAN, A.R.; GROVES-KIRKBY, C.J.; PHILLIPS, P. S.; CROCKETT, R.G.; WOOLRIDGE, A.; GILLMORE, G.K. The practical use of electrets in a public health radon remediation campaign, Journal of Environmental Radioactivity, v. 84(3), p. 375–391, 2005