Radiology and mammography standard X-ray spectra simulated with the Monte Carlo method
DOI:
https://doi.org/10.15392/bjrs.v9i2C.1670Palabras clave:
EGSnrc, TRS-457, HVLResumen
Six standard beams described in the TRS-457 (IAEA): RQR 5, 8, M1, M2, M3, M4 were simulated using the EGSnrc Monte Carlo code. Each spectrum was created by an X-ray tube simulated in BEAMnrc, and attenuation curves were obtained using the application egs_kerma. The quality of each beam was evaluated by the 1st and 2nd half-value layers, the homogeneity coefficients and the mean energies. All beams presented quality parameters compatible with those described in TRS-457 (IAEA).
Descargas
Referencias
IAEA - International Atomic Energy Agency. Dosimetry in diagnostic radiology: an international code of practice. Technical Reports Series No. 457. Vienna, 2007.
NAZÁRIO, M.; KHOURY, H. J.; HAZIN, C. Calibração de equipamentos de radioproteção com radiação gama no Laboratório de Metrologia das Radiações Ionizantes - DEN/UFPE. Metrologia 2003. Pernambuco, 2003.
PERINI, A. P.; NEVES, L. P.; SANTOS, W. S.; CALDAS, L. V. E. Experimental and Monte Carlo evaluation of an ionization chamber in a 60Co beam. Journal of Physics: Conference Series 733 012081, 2016.
HERRNSDORF, L.; ANDERSSON, M. ; GUNNARSSON, M. ; MATTSSON, S. Optimization of energy and directional response of a small 4π silicon dosemeter for quality control of CT/CBCT-units – A 3D CAD, Monte-Carlo, AM approach. Applied Radiation and Isotopes, v. 158 – 109050, p. 1 -12, 2020.
SALEHI, Z. ; BALVASI, E. ; AZIZ, M. Z. A. A review of the recent Monte Carlo (MC) simulation for dosimetry in mammographic applications. Malaysian Journal of Medicine and Health Sciences, v. 14(SUPP1), p. 76-81, 2018.
IAEA - International Atomic Energy Agency. Calibration of radiation protection monitoring instruments. Safety Reports Series No. 16. Vienna, 2000.
KAWRAKOW, I.; ROGERS, D. W. O. The EGSnrc code system: Monte Carlo simulation of electron and photon transport. National Research Council of Canada, NRCC. Report PIRS-701, 2003.
ROGERS, D. W. O.; WALTERS, B.; KAWRAKOW, I. BEAMnrc User’s Manual. National Research Council of Canada, NRCC. Report PIRS-0509(A)revL, 2010.
KAWRAKOW, I.; MAINEGRA-HING, E.; TESSIER, F.; TOWNSON, R.; WALTERS, B. EGSnrc C++ class library. Report PIRS-898, 2019.
MA, C.-M. and ROGERS, D. W. O. BEAMDP User’s Manual. National Research Council of Canada, NRCC. Report PIRS-0509(E)revA, 2020.
YU, M. K.; SLOBODA, R. S.; MURRAY, B. Linear accelerator photon beam quality at off-axis points. Medical Physics, v. 24(2), p. 233–239, 1997.
MATHIEU, K. B.; KAPPADATH S. C.; WHITE R. A.; ATKINSON E. N.; CODY D. D. An empirical model of diagnostic x-ray attenuation under narrow-beam geometry. Medical Physics, v. 38(8), p. 4546-4555, 2011.
ARNS, A. M. Towards clinical implementation of ultrafast combined kV-MV cone-beam CT for IGRT of lung tumors within breath-hold: evaluation of dosimetry and registration accuracy based on phantom studies. Doctorate dissertation – Heidelberg University, Heidelberg, 2016.
BAUR, M.; UHLMANN, N.; PÖSCHEL, T.; SCHRÖTER, M. Correction of beam hardening in X-ray radiograms. Review of Scientific Instruments, v. 90, 025108, 2019.
Descargas
Publicado
Número
Sección
Licencia
Derechos de autor 2021 Brazilian Journal of Radiation Sciences (BJRS)

Esta obra está bajo una licencia internacional Creative Commons Atribución 4.0.
Licencia: los artículos de BJRS tienen una licencia internacional Creative Commons Attribution 4.0, que permite el uso, el intercambio, la adaptación, la distribución y la reproducción en cualquier medio o formato, siempre que se otorgue el crédito correspondiente al autor o autores originales y a la fuente, proporcione un enlace a la licencia Creative Commons e indique si se realizaron cambios. Las imágenes u otros materiales de terceros en el artículo están incluidos en la licencia Creative Commons del artículo, a menos que se indique lo contrario en una línea de crédito al material. Si el material no está incluido en la licencia Creative Commons del artículo y su uso previsto no está permitido por la regulación legal o excede el uso permitido, el autor deberá obtener el permiso directamente del titular de los derechos de autor. Para ver una copia de esta licencia, visite http://creativecommons.org/licenses/by/4.0/