Acceptance testing of a CT scanner with a knowledge-based iterative reconstruction algorithm
DOI:
https://doi.org/10.15392/bjrs.v7i1A.456Palavras-chave:
Acceptance, CT, Iterative reconstructionResumo
Implementation of efficient computed tomography (CT) detectors, automatic exposure control and iterative reconstruction algorithms allows to obtain high quality images while minimizing absorbed dose to the patients, in consonance with the ALARA criterion. In this work, we present the results of the acceptance testing of a Philips iCT Elite scanner with the IMR reconstruction algorithm. A Catphan 500 phantom was used for this purpose. RESULTS: The laser positioning system was misaligned, while the gantry and table were aligned. The high-contrast spatial resolution was 21 lp/cm. Regarding low contrast resolution, a 3 mm pin with 0.3% contrast was observed using IMR for CTDIvol = 10.4 mGy, while a 5mm pin was observed using filtered back projection (FBP). Average uniformity was ± 3 Hounsfield Units (HU) and noise was <0.5%. It was verified that when using IMR reconstruction, the noise remained constant when the CTDIvol was reduced. Nominal and measured slice thicknesses were compared, resulting in the following values (measured/nominal): 4.82/5, 2.52/2.5, 1.24/1, 1.09/0.8. At 120 kVp, the effective energy was 70 keV. CT numbers were measured for different materials, with results in accordance with expected values. The contrast scale value resulted in 2·10-4 cm-1HU-1. CONCLUSIONS: Image quality and general system’s functionality were evaluated, with satisfactory results. The results were taken as a reference for routine quality controls. The IMR reconstruction algorithm resulted in significantly improved low-contrast resolution and lower noise at low doses when compared with FBP. The impact of IMR technology on clinical images remains to be assessed.
Downloads
Referências
SEERAM, E., Computed Tomography: Physical Principles, Clinical Applications, and Quality Control, 4th Edition. Philadelphia, EE.UU, 2016.
SCIBELLI, A., iDose4 iterative reconstruction technique, Philips Healthcare White Paper, 2011.
MEHTA, D; THOMPSON, R; MORTON, T; DHANANTWARI, A; SHEFER, E. Iterative model reconstruction: simultaneously lowered computed tomography radiation dose and improved image quality. Med Phys Int J, v. 1, p. 147-155, 2013.
Philips Healthcare. Philips iCT Elite Specifications, 2014.
IAEA. Human Health Series, Quality Assurance Programme for Computed Tomography: Diagnostic and Therapy Applications. Vienna, 2012.
ACR. Computed Tomography, Quality Control Manual, 2012.
AAPM – American Association of physicists in Medicine. Specification and acceptance testing of computed tomography scanners. AAPM Report 39, New York, 1993.
The Phantom Laboratory, Catphan 500 and 600 Manual. Salem, NY 2013.
SEFM – Sociedad Español de Física Médica. Protocolo Español de Control de Calidad en Radiodiagnóstico, Madrid: SEFM, 2011.
Downloads
Publicado
Edição
Seção
Licença
Direitos autorais (c) 2021 Brazilian Journal of Radiation Sciences

Este trabalho está licenciado sob uma licença Creative Commons Attribution 4.0 International License.
Declaro que o presente artigo é original, não tendo sido submetido à publicação em qualquer outro periódico nacional ou internacional, quer seja em parte ou em sua totalidade. Declaro, ainda, que uma vez publicado na revista Brazilian Journal of Radiation Sciences, editada pela Sociedade Brasileira de Proteção Radiológica, o mesmo jamais será submetido por mim ou por qualquer um dos demais co-autores a qualquer outro periódico. Através deste instrumento, em meu nome e em nome dos demais co-autores, porventura existentes, cedo os direitos autorais do referido artigo à Sociedade Brasileira de Proteção Radiológica, que está autorizada a publicá-lo em meio impresso, digital, ou outro existente, sem retribuição financeira para os autores.
Licença
Os artigos do BJRS são licenciados sob uma Creative Commons Atribuição 4.0 Licença Internacional, que permite o uso, compartilhamento, adaptação, distribuição e reprodução em qualquer meio ou formato, desde que você dê o devido crédito ao (s) autor (es) original (is) e à fonte, forneça um link para a licença Creative Commons, e indique se mudanças foram feitas. As imagens ou outro material de terceiros neste artigo estão incluídos na licença Creative Commons do artigo, a menos que indicado de outra forma em uma linha de crédito para o material. Se o material não estiver incluído no licença Creative Commons do artigo e seu uso pretendido não é permitido por regulamentação legal ou excede o uso permitido, você precisará obter permissão diretamente do detentor dos direitos autorais. Para visualizar uma cópia desta licença, visite http://creativecommons.org/licenses/by/4.0/