Acceptance testing of a CT scanner with a knowledge-based iterative reconstruction algorithm
DOI:
https://doi.org/10.15392/bjrs.v7i1A.456Keywords:
Acceptance, CT, Iterative reconstructionAbstract
Implementation of efficient computed tomography (CT) detectors, automatic exposure control and iterative reconstruction algorithms allows to obtain high quality images while minimizing absorbed dose to the patients, in consonance with the ALARA criterion. In this work, we present the results of the acceptance testing of a Philips iCT Elite scanner with the IMR reconstruction algorithm. A Catphan 500 phantom was used for this purpose. RESULTS: The laser positioning system was misaligned, while the gantry and table were aligned. The high-contrast spatial resolution was 21 lp/cm. Regarding low contrast resolution, a 3 mm pin with 0.3% contrast was observed using IMR for CTDIvol = 10.4 mGy, while a 5mm pin was observed using filtered back projection (FBP). Average uniformity was ± 3 Hounsfield Units (HU) and noise was <0.5%. It was verified that when using IMR reconstruction, the noise remained constant when the CTDIvol was reduced. Nominal and measured slice thicknesses were compared, resulting in the following values (measured/nominal): 4.82/5, 2.52/2.5, 1.24/1, 1.09/0.8. At 120 kVp, the effective energy was 70 keV. CT numbers were measured for different materials, with results in accordance with expected values. The contrast scale value resulted in 2·10-4 cm-1HU-1. CONCLUSIONS: Image quality and general system’s functionality were evaluated, with satisfactory results. The results were taken as a reference for routine quality controls. The IMR reconstruction algorithm resulted in significantly improved low-contrast resolution and lower noise at low doses when compared with FBP. The impact of IMR technology on clinical images remains to be assessed.
Downloads
References
SEERAM, E., Computed Tomography: Physical Principles, Clinical Applications, and Quality Control, 4th Edition. Philadelphia, EE.UU, 2016.
SCIBELLI, A., iDose4 iterative reconstruction technique, Philips Healthcare White Paper, 2011.
MEHTA, D; THOMPSON, R; MORTON, T; DHANANTWARI, A; SHEFER, E. Iterative model reconstruction: simultaneously lowered computed tomography radiation dose and improved image quality. Med Phys Int J, v. 1, p. 147-155, 2013.
Philips Healthcare. Philips iCT Elite Specifications, 2014.
IAEA. Human Health Series, Quality Assurance Programme for Computed Tomography: Diagnostic and Therapy Applications. Vienna, 2012.
ACR. Computed Tomography, Quality Control Manual, 2012.
AAPM – American Association of physicists in Medicine. Specification and acceptance testing of computed tomography scanners. AAPM Report 39, New York, 1993.
The Phantom Laboratory, Catphan 500 and 600 Manual. Salem, NY 2013.
SEFM – Sociedad Español de Física Médica. Protocolo Español de Control de Calidad en Radiodiagnóstico, Madrid: SEFM, 2011.
Published
Issue
Section
License
Copyright (c) 2021 Brazilian Journal of Radiation Sciences
This work is licensed under a Creative Commons Attribution 4.0 International License.
Licensing: The BJRS articles are licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/