Transit dose measurements using alanine and diode-based dosimeters

Authors

  • J. A. C. Gonçalves Instituto de Pesquisas Energéticas e Nucleares
  • E. S. R. Somessari Instituto de Pesquisas Energéticas e Nucleares
  • S. L. Somessari Instituto de Pesquisas Energéticas e Nucleares
  • Carmen C. Bueno Instituto de Pesquisas Energéticas e Nucleares - IPEN-CNEN/SP

DOI:

https://doi.org/10.15392/bjrs.v10i2A.1828

Keywords:

Si diode, transit dose, Gamma radiation processing, high dose dosimetry

Abstract

The growing interest in low-dose (< 100 Gy) radiation processing applications has raised concerns about accurately measuring the absorbed dose in irradiated materials. Depending on the irradiator design, the transit time due to the radioactive source movement (or the product itself) until the stable irradiation position might affect the predicted absorbed dose. This work aims to evaluate the transit dose in a 60Co Gammacell 220-Nordion irradiator, which has radioactive sources settled at the bottom of a lead shielding. When the facility is on, the product and the dosimeter are mechanically guided down to the irradiation position, and hereafter the selected exposure time starts to be counted. At the end of irradiation, both product and dosimeter rise to the initial position enabling them to be gathered by the operator. The product is continuously irradiated at different dose rates during its fall and rise movement, preventing the transit dose from being obtained straightforward. The experimental approach adopted is to assess the transit time, and thus the transit dose, using an online diode-based dosimetry system previously calibrated against reference standard alanine dosimeters. The agreement between the transit doses attained with the diode (0.41 ± 0.02) Gy and alanine ((0.38 ± 0.01) Gy validates the method herein proposed.

Downloads

Download data is not yet available.

References

ISO/ASTM - International Organization for Standardization/American Society for Testing and Materials. Standard Practice for Dosimetry in a Gamma Facility for Radiation Processing, ISO/ASTM 51702:2013, 3rd ed., 2013, 8p.

MANNAI, K., ASKRI, B., LOUSSAIEF, A., TRABELSI, A. Evaluation using Geant4 of the transit dose in the Tunisian gamma irradiator for insect sterilization. Applied Radiation and Isotopes, v.65, p. 701-707, 2007. DOI: https://doi.org/10.1016/j.apradiso.2007.01.011

LIN, M-H., CHAO, T-C., LEE, C-C., TUNG, C-J., YEH, C-Y., HONG, J-H. Measurement-based Monte Carlo dose calculation system for IMRT pretreatment and online transit dose verifications. Med Phys, v. 36(4), p. 1167-1175, 2009. DOI: https://doi.org/10.1118/1.3089790

FRONCOIS, P., BOISSARD, P., BERGER, L., MAZAL, A. In vivo dose verification from back-projection of a transit dose measurement on the central axis of photon beams. Physica Medica, v. 27, p. 1-10, 2011. DOI: https://doi.org/10.1016/j.ejmp.2010.06.002

BASTIN, K. T., PODGORSAK, M. B., THOMADSEN, B. R. The transit dose component of high dose rate brachytherapy: direct measurements and clinical implications. Int J Radiation Oncology Biol Phys, v. 26, p. 695-702, 1993. DOI: https://doi.org/10.1016/0360-3016(93)90291-3

FONSECA, G. P., RUBO, R. A., MINAMISAWA, R. A., DOS SANTOS, G. R., ANTUNES, P. C. G., YORIYAZ, H. Determination of transit dose profile for a 192Ir HDR source. Med Phys, v. 40, p. 051717, 2013. DOI: https://doi.org/10.1118/1.4802731

GONÇALVES, J. A. C.; MANGIAROTTI, A.; BUENO, C. C. Current response stability of a commercial PIN photodiode for low dose radiation processing applications. Rad Phys Chem, v. 167, p. 108276, 2020. DOI: https://doi.org/10.1016/j.radphyschem.2019.04.026

Downloads

Published

2022-07-21

How to Cite

A. C. Gonçalves, J., S. R. Somessari, E., L. Somessari, S., & Bueno, C. C. (2022). Transit dose measurements using alanine and diode-based dosimeters. Brazilian Journal of Radiation Sciences, 10(2A (Suppl.). https://doi.org/10.15392/bjrs.v10i2A.1828

Most read articles by the same author(s)