Feasibility studies of using thin entrance window photodiodes for clinical electron beam dosimetry

Authors

  • Cristina R. Nascimento Universidade Federal de Pernambuco - UFPE; Instituto Federal de Educação, Ciência e Tecnologia de Pernambuco – IFPE
  • Viviane K. Asfora Universidade Federal de Pernambuco - UFPE
  • Vinicius R. M. Barros Universidade Federal de Pernambuco - UFPE
  • Josemary A. C. Gonçalves Instituto de Pesquisas Energéticas e Nucleares - IPEN-CNEN/SP
  • Luiza F. R. Andrade Santa Casa de Misericórdia de Itabuna
  • Helen J. Khoury Universidade Federal de Pernambuco - UFPE
  • Carmen C. Bueno Instituto de Pesquisas Energéticas e Nucleares - IPEN-CNEN/SP

DOI:

https://doi.org/10.15392/bjrs.v7i2A.577

Keywords:

Electron dosimetry, Si photodiodes dosimeters, Solid-state dosimetry

Abstract

The response of the commercial XRA-24 PIN photodiode (5.76 mm2 active area) for clinical electron beam dosimetry covering the range of 8-12 MeV was investigated. Within this energy range, the charge generated in the diode’s sensitive volume is linearly dependent on the absorbed dose up to 320 cGy.  However, charge sensitivity coefficients evidenced that the dose response of the diode is slightly dependent on the electron beam energy. Indeed, the diode’s energy dependence was within 8.5% for 8-12MeV electron beams. On the other hand, it was also observed an excellent repeatability of these results with a variation coefficient (VC) lower than 0.4%, which is within the 1% tolerance limit recommended by the AAPM TG-62. Furthermore, the agreement between the percentage depth dose profiles (PDD) gathered with the diode and the ionization chamber allowed achieving the electron beam quality within 1% of that obtained with the ionization chamber. Based on these results, the photodiode XRA-24 can be a reliable and inexpensive alternative for electron beams dosimetry.

Downloads

Download data is not yet available.

Author Biographies

Cristina R. Nascimento, Universidade Federal de Pernambuco - UFPE; Instituto Federal de Educação, Ciência e Tecnologia de Pernambuco – IFPE

Departamento de Energia Nuclear - UFPE

Viviane K. Asfora, Universidade Federal de Pernambuco - UFPE

Departamento de Energia Nuclear

Vinicius R. M. Barros, Universidade Federal de Pernambuco - UFPE

Departamento de Energia Nuclear

Josemary A. C. Gonçalves, Instituto de Pesquisas Energéticas e Nucleares - IPEN-CNEN/SP

Centro de Tecnologia das Radiações

Helen J. Khoury, Universidade Federal de Pernambuco - UFPE

Departamento de Energia Nuclear

Carmen C. Bueno, Instituto de Pesquisas Energéticas e Nucleares - IPEN-CNEN/SP

Centro de Tecnologia das Radiações

References

GIESSBACH, I.; LAPP, M.; BOHSUNG, J.; GADEMANN, G.; HARDE, D. Dosimetric characteristics of a new unshield silicon diode and its application in clinical photon and electron beams. Medical Physics, v. 32, p. 12, 2005.

ROSENFELD, A. B.; CUTAJAR, D.; LERCH, M. L. F.; TAKACS, G.; CORNELIUS, I. M. ; YUDELEV, M.; ZAIDER, M. Miniature semiconductor detectors for in vivo dosimetry. Radiation Protection Dosimetry, v. 120 (1-4), p. 48-55, 2006.

SONG, H.; AHMAD, M.; DENG, J.; CHEN, Z.; YUE, N. J.; NATH, R. Limitations of silicon diodes for clinical electron dosimetry. Radiation Protection Dosimetry, v. 120 (1-4), p. 56-59, 2006.

SANTOS, T. C.; NEVES, W. F. P.; GONÇALVES, J. A. C.; HADDAD, C. M. K.; HARKONEN, J.; BUENO, C. C. Characterization of miniature RAD-HARD silicon diodes as dosimeters for small fields of photon beams used in radiotherapy. Radiation Measurements, v. 71, p. 396-401, 2014.

GONÇALVES, J. A. C.; PEREIRA, L. N.; POTIENS, M. P. A.; VIVOLO, V.; BUENO, C. C. Evaluation of epitaxial silicon diodes as dosimeters in X-ray mammography. Radiation Measurements, v. 71, p. 384-388, 2014.

BRUZZI, M.; BUCCIOLINI, M.; CASATI, M.; MENICHELLI, D.; TALAMONTI, C.; PIEMEONTE, C.; SVENSSON, B. G. Epitaxial silicon devices for dosimetry applications. Applied Physics Letters, v. 90, p. 172109, 2007.

IAEA. Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water. International Atomic Energy Agency Technical Reports Series (IAEA-TRS) n. 398. Vienna, 2000.

IAEA. The Use of Plane-Parallel Ionization Chambers in High-Energy and Photons Beams: An International Code of Practice for Dosimetry. International Atomic Energy Agency Technical Reports Series (IAEA-TRS) n. 381. Vienna, 1997.

AAPM. Diode in vivo dosimetry for patients receiving external beam radiation therapy. Report of American Association of Physicists in Medicine Radiation Therapy Committee Task Group 62. AAPM- TG-62 Report 87. New York: AAPM, 2005.

SORIANI, A.; DE ANGELIS, C.; ONORI, S.; IACCARINO, G.; MARZI, S.; LANDONI, V.; BENASSI, M. Sistemi a stato sólido per la dosimetria di fasci di elettroni ad alto rateo di dose. In: III Congresso Nazionale dell'Associazione italiana di Fisica in Medicina (AIFM), Agrigento, 2003.

KHOURY, H. J.; HAZIN, C. A.; MASCARENHAS, A. P.; SILVA Jr, E. F. Low cost silicon photodiode for electron dosimetry. Radiation Protection Dosimetry, v. 84 (1-4), p. 341-343, 1999.

Downloads

Published

2019-02-07

How to Cite

Nascimento, C. R., Asfora, V. K., Barros, V. R. M., Gonçalves, J. A. C., Andrade, L. F. R., Khoury, H. J., & Bueno, C. C. (2019). Feasibility studies of using thin entrance window photodiodes for clinical electron beam dosimetry. Brazilian Journal of Radiation Sciences, 7(2A (Suppl.). https://doi.org/10.15392/bjrs.v7i2A.577

Issue

Section

The Meeting on Nuclear Applications (ENAN)