Methodology for preparing a reference sample in the air filter matrix

Authors

  • Alessandra dos Santos Fernandes
  • Poliana Santos de Souza
  • Estela Maria Oliveira IRD-CNEN
  • Johnny de Almeida Rangel
  • Wanderson de Oliveira Sousa
  • Ricardo Tadeu Lopes

DOI:

https://doi.org/10.15392/bjrs.v10i2.2037

Keywords:

Reference material, Air Filter, Radionuclide Metrology, Pantograph

Abstract

The proficiency tests are widely used to evaluate the analytical capacity of laboratories and also as part of the accreditation processes. The National Intercomparison Program is a Proficiency Test that uses radioactive    reference materials in soil, vegetation, water and air filters. This work is a study of the choice of composition of radionuclides ideal to be used in the air filters to be produced by LNMRI as reference material for PNI.

Five cocktails composed of different radionuclides chosen for their energies in the range of 30 to 2000 keV were prepared, each cocktail was distributed in three glass fiber air filters using a pantograph using the pycnometer technique, contaminating each filter with 19 (nineteen) drops, totaling fifteen (15) air filter samples, with the objective of homogeneous distribution of the droplets at the sites determined by the geometry chosen without the need for marking in the filters, in order to choose the best composition to be used in the preparation of reference material.

Downloads

Download data is not yet available.

References

VIANNA, M. E.; TAUHATA, L.; CONCEIÇÃO, C. C. S.; OLIVEIRA, A.; OLIVEIRA, J.; CLAIN, A. F.; GARCIA, L. C. Quality of radionuclide analysis in environmental samples. The Science of Total Environmental, v. 173/174, p. 15-18, 1995. DOI: https://doi.org/10.1016/0048-9697(95)04745-X

VIANNA, M. E.; TAUHATA, L.; CLAIN, A. F.; FERREIRA, A.; OLIVEIRA, A.; OLIVEIRA, J. Analysis of Brazilian Intercomparison program data from 1991 to 1995 of radionuclide determination in environmental samples. Applied Radiation and Isotopes, v. 49, p. 1463-1466, 1998. DOI: https://doi.org/10.1016/S0969-8043(97)10093-8

TAUHATA, L.; OLIVEIRA, A.; OLIVEIRA, J.; CLAIN, A.F.; VIANNA, M.E.; FERREIRA, A. C. M.; BRAGANÇA, M. J. C.; DE FARIA, R. Q. The Brazilian National Intercomparison (PNI/IRD/CNEN): evaluation of 15 years of data. Journal of Environmental Radioactivity Applied, v. 86, p. 384-390, 2006. DOI: https://doi.org/10.1016/j.jenvrad.2005.10.002

DE FELICE, P. ; CECCATELLI, A. ; FAZIO, A. Development of simulated air filters gamma-ray spectrometry proficiency testing. Applied Radiation and Isotopes. v. 68, p. 1240-1246, 2010. DOI: https://doi.org/10.1016/j.apradiso.2010.01.010

CAMPION, P. J. Procedures for Accurately Diluting and Dispensing Radioactive Solutions. Bureau International des Poids et Mesures (Monograph BIPM-l), 1975

LOURENÇO, V.; BOBIN, C. Weighing uncertainties in quantitative source preparation for ra-dionuclide metrology. Metrologia. v. 52, p. S18–S29, 2015. DOI: https://doi.org/10.1088/0026-1394/52/3/S18

Downloads

Published

2022-06-13

How to Cite

Fernandes, A. dos S., de Souza, P. S., Oliveira, E. M., Rangel, J. de A., Sousa, W. de O., & Lopes, R. T. (2022). Methodology for preparing a reference sample in the air filter matrix. Brazilian Journal of Radiation Sciences, 10(2). https://doi.org/10.15392/bjrs.v10i2.2037

Issue

Section

Articles

Most read articles by the same author(s)

<< < 1 2