Development of alternative methodology for normalization of the small animals PET scanner of LIM/CDTN

Authors

DOI:

https://doi.org/10.15392/2319-0612.2024.2360

Keywords:

Normalization, Ge-68 source, 18F-FDG, Small Animal PET Scanner

Abstract

Normalization is a step taken to correct variations in count rates between detection channels of small animals PET scanners. In general, in laboratory practice, normalization is performed using a commercial source of germanium 68. Given the half-life of the Ge-68 isotope, the normalization source must be replaced every two years at most, which generates a significant spent for the Laboratory. The objective of this work was to develop and test an alternative rechargeable source to be used on normalization routine. The results indicated that the rechargeable source (18F-FDG) can be used to replace the commercial source.

Downloads

Download data is not yet available.

Author Biographies

Victória Moreira, Centro de Desenvolvimento da Tecnologia Nuclear CDTN/CNEN

UFMG-Radiologia

Rodrigo Gadelha, CNEN / UFMG / EBSERH – HC

UFMG-Departamento de Anatomia e Imagem

EBSERH

References

SAHA, G. B. Basics of PET imaging, Second Edition. 2. ed. Cleveland: Springer, 2010. 244 p. ISBN 978-1-4419-0804-9

GE Healthcare Technologies, Triumph Service Guide Technical Publication. Revision Draft 6, Copyright. Available in: <http://www.gehealthcare.com>. 2011.

QRM. Micro-PET IQ Phantom, NEMA NU4, for performance evaluation of Micro-PET systems according to NEMA NU 4 , available in www.qrm.de, consulted in 06/13/2023.

NEMA - National Electrical Manufacturers Association. Performance Measurements of Small Animal Positron Emission Tomographs. Rosslyn VA; 2008 Standards Publication NU 4-2008

GONTIJO, R. M. G. Image quality evaluation of a small animal PET scanner. Brazilian Journal of Radiation Sciences, 08-01(2020)

BELCARI, N. et al., NEMA NU-4 Performance Evaluation of the IRIS PET/CT Preclinical Scanner. IEEE Transactions on Nuclear Science. 2017. v. 1. p. 301-309. 2017. DOI: https://doi.org/10.1109/TRPMS.2017.2707300

Downloads

Published

2024-02-23

How to Cite

Silva, A. C. M. da, Moreira, V. G., Mendes, B. M., Gontijo, R. M. G., & Ferreira, A. V. (2024). Development of alternative methodology for normalization of the small animals PET scanner of LIM/CDTN. Brazilian Journal of Radiation Sciences, 12(1), e2360. https://doi.org/10.15392/2319-0612.2024.2360

Issue

Section

Articles

Most read articles by the same author(s)

<< < 1 2 3