Abstract

An alternative simple method to evaluate the L/D ratio of neutron imaging beams is proposed. It is based on the behavior of the ratio “x/Ut”, which asymptotically tends to L/D for large values of the parameter “x”, that corresponds to the distance separating, a gadolinium test object to the scintillator plane, where its image is formed. The method was applied to the neutron imaging equipment of the Nuclear and Energy Research Institute IPEN-CNEN/SP and the obtained result was L/D = (104 ± 4). The consistency of the proposed method was verified by comparing this value, with those ones obtained by a well - known and established procedure.

Downloads

Download data is not yet available.

References

American Society for Testing and Materials. Determining the L/D ratio of neutron radi-ography beams: ASTM, (E803), 1981.

VON DER HARDT, P.; ROETTGER, H. Neutron Radiography Handbook: Nuclear Science and Technology. Dordrecht, D.Reidl, 1981.

HARMS, A. A.; ZEILINGER, A. A. New Formulation of Total Unsharpness in Radiograph, Phys. Med. Biol., v. 22, no 1, p. 70–80. 1977.

WADE, ,J. R.; HOWARD, A. L. Experiments at Argonne National Laboratory, Nucl. Technol., v. 76, no 3, p. 408–419, 1987.

STANOJEV PEREIRA, M. A.; SCHOUERI, R.; DOMIENIKAN, C.; TOLEDO, F.; ANDRADE, M. L. G.; PUGLIESI, R. The neutron tomography facility of IPEN-CNEN/SP and its potential to investigate ceramic objects from the Brazilian cultural her-itage, Appl. Rad. and Isot., v. 75, p. 6-10, 2013.

SCHOUERI, R.; DOMIENIKAN, C.; TOLEDO, F.; ANDRADE, M. L. G.; STANOJEV PEREIRA, M. A.; PUGLIESI, R. The new facility for neutron tomography of IPEN-CNEN/SP and its potential to investigate hydrogenous substances, Appl. Rad. and Isot., v. 84, p. 22-26, 2014.

WROBEL, M.; GREIM, L. Resolution functions and unsharpness in neutron radiog-raphy. GKKS/e/12th ed. Geesthacht, Germany: GKKS, 1988.