Validating a virtual source model based in Monte Carlo method for profiles and percent depth doses calculation

Authors

  • Renata Aline Del Nero Instituto de Pesquisas Energeticas e Nucleares (IPEN / CNEN - SP)
  • Marcos Vinicius Nakaoka Nakandakari Hospital Beneficencia Portuguesa de Sao Paulo
  • Hélio Yoriyaz Instituto de Pesquisas Energeticas e Nucleares (IPEN / CNEN - SP)

DOI:

https://doi.org/10.15392/bjrs.v7i2A.689

Keywords:

(virtual source model, Monte Carlo method, radiotherapy)

Abstract

The Monte Carlo method for radiation transport has been adapted for medical physics application. More specifically, it has received more attention in clinical treatment planning with the development of more efficient computer simulation techniques. In linear accelerator modeling by the Monte Carlo method, the phase space data file (phsp) is an alternative representation for radiation source. However, to create a phase space file and obtain good precision in the results, it is necessary detailed information about the accelerator's head and commonly the supplier does not provide all the necessary data. An alternative to the phsp is the Virtual Source Model (VSM). This alternative approach presents many advantages for the clinical Monte Carlo application. This is the most efficient method for particle generation and can provide an accuracy similar when the phsp is used. This research propose a VSM simulation with the use of a Virtual Flattening Filter (VFF) for profiles and percent depth doses calculation. Two different sizes of open fields (40 x 40 cm² and 40 x 40 cm² rotated 45°) were used and two different source to surface distance (SSD) were applied: the standard 100 cm and custom SSD of 370 cm, which is applied in radiotherapy treatments of total body irradiation. The data generated by the simulation was analyzed and compared with experimental data to validate the VSM. This current model is easy to build and test.

Downloads

Download data is not yet available.

References

ANDREO, P. Monte Carlo techniques in medical radiation physics. Phys. Med. Biol., v. 36, n. 7, p. 861-920, 1991.

CHAVES, A.; LOPES, M.C.; ALVES, C.C.; OLIVERA, C.; PERALTA, L.; RODRI-GUES, P.; TRINDADE, A. A Monte Carlo multiple source model applied to radiosurgery nar-row photon beams. Medical Physics, v. 31, n. 8, p. 2192-2204, 2004.

AALTONEN, P.; BRAHME, A.; LAX, I.; LEVERNES, S.; NASLUND, I.; REITAN, J.B.; TURESSON, I. Specification of dose delivery in radiation therapy. Acta Oncologica, v. 36, p. 1-32, 1997.

ICRU, International Commission on Radiation Units and Measurements. ‘‘Prescribing, Re-cording and Reporting Photon Beam Therapy’’, Report No. 62, Supplement to ICRU Report No. 50, Bethesda, MD, CRU, 1999.

PLESSIS, F.C.P. DU; WILLEMSE, C.A.; LO¨TTER, M.G. Comparison of Batho, ETAR and Monte Carlo dose calculation methods in CT based patients models. Medical Physics, v. 28, p. 582-589, 2001.

SIKORA, M.; ALBER, M. A virtual source model of electron contamination of a thera-peutic photon beam. Phys. Med. Biol., v. 54, n. 24, p. 7329-7344, 2009.

MA, C.M.; FADDEGON, B.A.; ROGERS, D.W.O.; MACKIE, T.R. Accurate characteri-zation of Monte Carlo calculated electron beams for radiotherapy. Medical Physics, v. 24, n. 3, p. 401-416, 1997.

YANG, J.; LI, J.S.; QUIN, L.; XIONG, W.; MA, C.M. Modelling of electron contamina-tion in clinical photon beams for Monte Carlo dose calculation. Phys. Med. Biol., v. 49, n. 12, p. 2657-2673, 2004.

VERHAEGEN, F.; SEUNTJENS, J. Monte Carlo modeling of external radiotherapy photon beams. Phys. Med. Biol., v. 48, n. 21, p. 107-164, 2003.

CAPOTE, R.; JERAJ, R.; MA, C.M.; ROGERS, D.W.O.; SÁNCHEZ-DOBLADO, F.; SEMPAU, J.; SEUNTJENS, J.; SIEBERS, J. V. Phase-space database for external beam radio-therapy. Summary report of a consultants’ meeting. Vienna: International Atomic Energy Agency; 2006.

MA, C.M. Characterization of computer simulated radiotherapy beams for Monte Car-lo treatment planning. Radiation Physics Chemistry, v. 53, p. 329-344, 1998.

SIKORA, M.; DOHM, O.; ALBER, M. A virtual photon source model of an Elekta line-ar accelerator with integrated MLC for Monte Carlo based dose calculation. Phys. Med. Biol., v. 52, n. 15, p. 4449-4463, 2007.

RUCCI, A.; CARLETTI, C.; CRAVERO, W.; STRBAC, B. Use of IAEA's phase-space files for the implementation of a clinical accelerator virtual source model. Physica Medica, v. 30, p. 242-248, 2014.

RUCCI, A.; CARLETTI, C.; CRAVERO, W.; STRBAC, B. Use of IAEA's phase-space files for virtual source model implementation: Extension to large fields. Physica Medica, v. 32, p. 1030-1033, 2016.

ANDREO, P.; BURNS, D. T.; HOHLFELD, K.; HUQ, M. S.; KANAI, T.; LAITANO, F.; SMYTH, V.; VYNCKIER, S. Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water. International Atomic Energy Agency TRS-398, Vienna, 2000.

GOORLEY, J. T.; JAMES, M. R.; BOOTH, T. E.; BROWN, F. B.; BULL, J. S.; COX, L. J.; DURKEE, J. W.; ELSON, J. S.; FENSIN, M. L.; FORSTER, R. A.; HENDRICKS, J. S.; HUGHES, H. G.; JOHNS, R. C.; KIEDROWSKI, B. C.; MARTZ, R. L.; MASHNIK, S. G.; McKINNEY, G. W.; PELOWITZ, D. B.; PRAEL, R. E.; SWEEZY, J. E.; WATERS, L. S.; WIL-COX, T. A.; ZUKAITIS, A. MCNP6TM User’s Manual - Version 1.0. LA-CP-13-00634, Rev. 0. Los Alamos National Security, LLC. 2013.

PÖNISCH, F.; TITT, U.; VASSILIEV, O.N.; KRY, S.F.; MOHAN, R. Properties of un-flatted photon beams shaped by a multileaf collimator. Medical Physics, v. 33, n. 6, p. 1738-1746, 2006.

JAVEDAN, K.; FEYGELMAN, V.; ZHANG, R.R.; MOROS, E.G.; CORREA, C.R.; TROTTI, A.; LI, W.; ZHANG, G.G. Monte Carlo comparison of superficial dose between flat-tening filter free and flattened beams. Physica Medica, v. 30, n. 4, p. 503-508, 2014.

BUTSON, M.J.; MATHUR, J.N.; METCALFE, P.E. Skin dose from radiotherapy X-ray beams: the influence of energy. Australasian Radioly, v. 41, n. 2, p. 148-150, 1997.

Downloads

Published

2019-02-20

How to Cite

Del Nero, R. A., Nakaoka Nakandakari, M. V., & Yoriyaz, H. (2019). Validating a virtual source model based in Monte Carlo method for profiles and percent depth doses calculation. Brazilian Journal of Radiation Sciences, 7(2A (Suppl.). https://doi.org/10.15392/bjrs.v7i2A.689

Similar Articles

You may also start an advanced similarity search for this article.

Most read articles by the same author(s)