RADIOCHROMIC FILM CALIBRATION FOR DOSIMETRY IN CHEST CT SCANS

Authors

  • Mauricio Gomez Universidade Federal de Minas Gerais
  • Arnaldo Prata Mourão Universidade Federal de Minas Gerais

DOI:

https://doi.org/10.15392/bjrs.v9i1A.1284

Keywords:

Computed Tomography, Radiochromic film, Calibration

Abstract

Phantoms can be used as quality control objects in computed tomography (CT) imaging and dosimetry in specific regions of interest. By means of phantom tests, alternatives can be created to optimize the acquisition protocols and thus modify the acquisition parameters to deposit lower levels of radiation to the patient, seeking a quality image for the diagnosis. Radiochromic films, GAFCHROMIC XR-AQ2 model, are used to measure the dose-response when exposed to radiation. The profile of the absorbed dose within a phantom is important because it allows an understanding of a larger reality. Calibration curves were obtained from radiochromic films using a cylindrical phantom made of polymethyl methacrylate (PMMA). This phantom has five openings, one central and four peripherals lagged 90°. The phantom was positioned in the isocenter of the gantry of a scanner manufactured by the Toshiba, model Asterion, using 120 and 135 kV voltage of the X-ray tube. The phantom peripheral openings were positioned similarly to positions 3, 6, 9, and 12 of the analog clock. The ionization chamber was used for the record values of the Air Kerma in PMMA, in the central and 12 positions when the central slice of the phantom was irradiated in the axial mode. The other openings were filled with rods also made of PMMA. Radiochromic film strips were placed into PMMA rods and the central and 12 openings were filled. The central area of the phantom was scanned in helical mode. The film strips were digitized and processed their images to have the calibration curve survey.

Downloads

Download data is not yet available.

References

MOURÃO, A.P AND OLIVEIRA, F.A, Fundamentos de radiologia e imagem. São Paulo: Difusão Editora. First edition, 2009.

DOVALES, A. C. ET AL. Computed tomography in Brazil: frequency and pattern of usage among inpatients of the unified health system (sus). Revista Brasileira de Fisica Medica, v. 9, n. 1, p. 11-14, 2015.

DOVALES, A. C. ET AL. Patterns and trends of computed tomography usage in outpatients of the Brazilian public healthcare system, 2001-2011 Journal of Radiological Protection, IOP Publishing, v. 36, n. 3, p. 547, 2016.

GREFFIER, J. ET AL. Dose reduction with iterative reconstruction: optimization of ct protocols in clinical practice. Diagnostic and interventional imaging, Elsevier, v. 96, n. 5, p. 477-486, 2015.

GABUSI, M. ET AL. Radiation dose in chest ct: assessment of size-specific dose estimates based on water-equivalent correction. Physica Medica. Elsevier, v. 32, n. 2, p. 393-397, 2016.

GOMEZ, A. Estudo de dosimetria e qualidade de imagem em varreduras de tomografia computadorizada de cabeça utilizando objeto simulador. Dissertação (mestrado) Universidade Federal de Minas Gerais, Escola de Engenharia, 2017.

XU, X. G.; ECKERMAN, K. F. Handbook of anatomical models of radiation dosimetry. [S.l]: CRC press, 2009.

OLIVEIRA, A. C DE. ET AL. Construção de fantomas físicos para avaliação dosimétrica em planejamentos radioterapêuticos: O fantoma hydra.

ASHLAND INC. GafchromicTM XR film. GafchromicTM Radiology Films, [internet] Access Dec 10th 2015. <http://www.ashland.com/products/gafchromic-radiology-films>.

RADCAL, C. ACCU-GOLD User Guide. [S.l.], 2015.

Downloads

Published

2021-04-30

Issue

Section

The Meeting on Nuclear Applications (ENAN) 2019

How to Cite

RADIOCHROMIC FILM CALIBRATION FOR DOSIMETRY IN CHEST CT SCANS. Brazilian Journal of Radiation Sciences, Rio de Janeiro, Brazil, v. 9, n. 1A, 2021. DOI: 10.15392/bjrs.v9i1A.1284. Disponível em: https://bjrs.org.br/revista/index.php/REVISTA/article/view/1284.. Acesso em: 24 nov. 2024.

Similar Articles

31-40 of 196

You may also start an advanced similarity search for this article.

Most read articles by the same author(s)