Image quality evaluation of a small animal PET scanner

Authors

  • Andréa Vidal Ferreira Centro de Desenvolvimento da Tecnologia Nuclear CDTN/CNEN
  • Rodrigo Modesto Gadelha Gontijo Faculdade de Medicina da UFMG
  • Guilherme Cavalcante de Albuquerque Souza Centro de Desenvolvimento da Tecnologia Nuclear CDTN/CNEN
  • Bruno Melo Mendes Centro de Desenvolvimento da Tecnologia Nuclear CDTN/CNEN
  • Juliana Batista da Silva
  • Marcelo Mamede Faculdade de medicina da UFMG

DOI:

https://doi.org/10.15392/bjrs.v8i1.1155

Keywords:

Small Animal PET Scanner, Image Quality, NEMA NU 4-2008

Abstract

Performance testing of small animal PET scanners is very important to ensure the high performance required for this particular type of PET scanner. In this sense, National Electrical Manufactures Association (NEMA) published its NU 4/2008 standards, a consistent and standardized methodology for measuring scanner performance parameters for small animal PET imaging. Imaging capabilities of the scanner LabPET SOLO 4 were tested during two years according NEMA NU 4/2008 methodology. Results indicates that the equipment, despite a decade of use, presents performance appropriated and similar to performance of equipments reported in Literature. 

Downloads

Download data is not yet available.

References

YAO R.; LECOMTE R.; CRAWFORD E. Small-Animal PET: What is it, and why do we need it? Journal of Nuclear Medicine Technology. 2011. v.40 nº3, p.157-165. DOI: https://doi.org/10.2967/jnmt.111.098632

MANNHEIM JG et al., Reproducibility and comparability of preclinical PET imaging data: A multi-center small animal PET study Journal of Nuclear Medicine. Published on March 8, 2019 as doi:10.2967/jnumed.118.221994. DOI: https://doi.org/10.2967/jnumed.118.221994

BAO et al., Performance evaluation of the Inveon dedicated PET preclinical tomograph based on the NEMA NU-4 standards. Journal of Nuclear Medicine. 2009. 50:401-408. DOI: https://doi.org/10.2967/jnumed.108.056374

NEMA - National Electrical Manufacturers Association. Performance Measurements of SmallAnimal Positron Emission Tomographs. Rosslyn VA; 2008 Standards Publication NU 4-2008.

NEMA - National Electrical Manufacturers Association. Performance Measurements of Positron Emission Tomographs (PETS). Rosslyn VA; Standards Publication NU 2-1994.

NEMA - National Electrical Manufacturers Association. Performance Measurements of Positron Emission Tomographs (PETS). Rosslyn VA; Standards Publication NU 2-2001.

NEMA - National Electrical Manufacturers Association. Performance Measurements of Positron Emission Tomographs (PETS). Rosslyn VA; Standards Publication NU 2-2007.

NEMA - National Electrical Manufacturers Association. Performance Measurements of Positron Emission Tomographs (PETS). Rosslyn VA; Standards Publication NU 2-2018.

GE Healthcare Technologies. Triumph Service Guide Technical Publication. Revision Draft 6, Copyright. Available at: <http://www.gehealthcare.com>. 2011.

TRETRAULT M. A., et al. System architecture of the LabPET small animal PET scanner. IEEE Transactions on Nuclear Science, 2008. v.55, p. 2546-2550. DOI: https://doi.org/10.1109/TNS.2008.2002326

FONTAINE R., et al. The hardware and signal processing architecture of LabPETTM, a small animal APD-based digital PET scanner. IEEE Transactions on Nuclear Science, February 2009. v. 56, p. 3-9. DOI: https://doi.org/10.1109/TNS.2008.2007485

Gamma Medica-Ideas, User Manual: Positron Emission Tomography (PET) Scanner Software - Manual for Acquisition and Processing of PET Images from the LABPET System. Software version: LABPET 1.12.0, (2010).

BELCARI, N. et al. NEMA NU-4 Performance Evaluation of the IRIS PET/CT Preclincial Scanner. IEEE Transactions on Nuclear Science, 2017. v.1, n.4, p. 301-309. DOI: https://doi.org/10.1109/TRPMS.2017.2707300

PMOD Tecnologies LCC, PMOD v.3.7. User Manual, (2015).

PRASAD, R., RATIB, O., ZAID, H., NEMA NU-04-based performance characteristics of the LabPET-8TM small animal PET scanner, Phys Med Biol. 2011, 56(20):6649-6664. DOI: https://doi.org/10.1088/0031-9155/56/20/009

GOERTZEN et al. NEMA NU 4-2008 Comparison of Preclinical PET Imaging Systems. 2012, JNM, v. 56, p. 1300-1309. DOI: https://doi.org/10.2967/jnumed.111.099382

Downloads

Published

2020-03-22

How to Cite

Ferreira, A. V., Gontijo, R. M. G., Souza, G. C. de A., Mendes, B. M., Silva, J. B. da, & Mamede, M. (2020). Image quality evaluation of a small animal PET scanner. Brazilian Journal of Radiation Sciences, 8(1). https://doi.org/10.15392/bjrs.v8i1.1155

Issue