Detection limit calculation according to ISO 11929 for in vitro 210Pb radiobioassay determinations by LSC

Authors

DOI:

https://doi.org/10.15392/2319-0612.2023.2185

Keywords:

detection limit, 210Pb, LSC

Abstract

According to the limit of detection (DL) presented in ISO 11929:2010, the International Organization for Standardization (ISO) abandoned the concept of minimum detectable amount (MDA) described in ISO 12790-1:2001. In this work we demonstrate and discuss how we calculate the detection limit for the determination of 210Pb by liquid scintillation counter (LSC) for the previously developed methodology and compare its results with the MDA value. The DL value found was 0.033 Bq/L instead of 0.032 Bq/L for MDA. The MDA and LD values ​​did not differ virtually, the results highlight that the main change between the two Standard is related to the definition approach.

Downloads

Download data is not yet available.

References

ISO 12790-1:2001. Performance criteria for radiobioassay — Part 1: General principles. <https://www.iso.org/standard/32473.html>

ISO 11929:2010. Determination of the characteristic limits (decision threshold, detection limit and limits of the confidence interval) for measurements of ionizing radiation — Fundamentals and application. < https://www.iso.org/standard/43810.html>

ISO 28218:2010. Performance criteria for Radiobioassay. <https://www.iso.org/standard/40869.html>

SAMPAIO, C.S.; MEDEIROS, G.C.O.; MESQUITA, S.A.; DANTAS, B.M.; SOUSA, W.O. A new approach for the determination of 210Pb by liquid scintillation counting. Appl. Radiat. Isot. 156 108972, 2020. < https://doi.org/10.1016/j.apradiso.2019.108972> DOI: https://doi.org/10.1016/j.apradiso.2019.108972

Avaliação de dados de medição: Guia para a expressão de incerteza de medição – GUM 2008. Duque de Caxias, RJ: INMETRO/CICMA/SEPIN, 2012, 141 p. < https://www.gov.br/inmetro/pt-br/centrais-de-conteudo/publicacoes/documentos-tecnicos-em-metrologia/gum_final.pdf/view>

CURRIE, L. Limits for Qualitative Detection and Quantitative Determination. Anal. Chem. 40; pp. 586-593, 1968 < https://doi.org/10.1021/ac60259a007> DOI: https://doi.org/10.1021/ac60259a007

Downloads

Published

2023-03-29

How to Cite

Detection limit calculation according to ISO 11929 for in vitro 210Pb radiobioassay determinations by LSC. Brazilian Journal of Radiation Sciences, Rio de Janeiro, Brazil, v. 11, n. 1A (Suppl.), p. 1–9, 2023. DOI: 10.15392/2319-0612.2023.2185. Disponível em: https://bjrs.org.br/revista/index.php/REVISTA/article/view/2185.. Acesso em: 22 nov. 2024.

Similar Articles

81-90 of 136

You may also start an advanced similarity search for this article.