Characterization of a Commercial PIN Diode for Stereotactic Radiosurgery Dosimetry

Authors

  • Matheus Fernando dos santos Universidade Federal de Pernambuco https://orcid.org/0000-0003-0477-249X
  • Charles Nilton Do Prado Oliveira UFPE
  • Karen Pieri Bueno
  • Lucas Delbem Albino
  • Thiago Schmeling Fontana
  • Wlademir Carvalho Vasconcelos Filho
  • Jose Geneci da Costa Calado
  • Ernesto Henrique Roesler
  • Vinicius Saito Monteiro de Barros
  • Helen Jamil Khoury

DOI:

https://doi.org/10.15392/bjrs.v10i2A.1827

Keywords:

diode, small fields, dosimetry, circular cones. radiosurgery

Abstract

The stereotactic radiosurgery (SRS) is a technique that uses multiple beams extremely collimated of ionizing
radiation to treat intracranial lesions and functional abnormalities, with high geometric precision and dosimetric
accuracy. The use of small fields is already a reality in modern radiotherapy techniques. However, the accuracy
in small-field dosimetry is challenging because of several process physics and aspects related to the detector. The
aim of this paper is to evaluate the response of a dosimetric system developed by Nuclear Energy Department of
Federal University of Pernambuco (DEN/UFPE) for small field dosimetry. Measurements of, output factor (OF),
off-axis ratio (OAR) and percentage depth dose (PDP) were performed with this dosimetric system and
compared with the results obtained with a commercial diode and treatment planning system data. The results
showed that the Diode-DEN presents a reliable and economical alternative for small fields dosimetry used in
treatments of SRS. This detector can be used for validation of dates obtained in commissioning of linear
accelerator and inserted in Treatment Planning Systems (TPS).

Downloads

Download data is not yet available.

Author Biography

  • Matheus Fernando dos santos, Universidade Federal de Pernambuco
    physics

References

PARK S. Y.; CHOI B. G.; LEE D. M.; JANG N. Y. Analysis of small-field dosimetry with

various detectors. Korean Society of Medical Physics, Progress in Medical Physics, v. 29,

SCHMELING T.F. Validation of the frameless functional radiosurgery technique using cone for

trigeminal neuralgia cases (Dissertation) 2021.

IAEA – International Atomic Energy Agency. Technical Reports Series n° 483- Dosimetry of

small static fields used in external beam radiotherapy. An international code of practice for

reference and relative dose determination. IAEA, Viena, 2017.

KUMAR et al. Dosimetric Characteristics of a PIN Diode for Radiotherapy Application.

Technology in Cancer Research and Treatment. v.13 n. 4, 2014.

GALVAN J. Comissioning of integrated system of radiosurgery Varian ICVI. (Dissertation)

CHENG J. Y. Output factor comparison of Monte Carlo and measurement for Varian TrueBeam

MV and 10 MV flattening filter-free stereotactic radiosurgery system. Journal of Applied

Clinical Medical Physics, v. 17,n.3, p.100-110, 2018.

VENSELAAR J., WELLEWEERD H. ; MIJNHEER B. Tolerances for accuracy of photon

beam dose calculation of treatment planning systems. Radiotherapy and Oncology. V. 60, p.

-201, 2001.

Downloads

Published

2022-07-21

How to Cite

Characterization of a Commercial PIN Diode for Stereotactic Radiosurgery Dosimetry. Brazilian Journal of Radiation Sciences, Rio de Janeiro, Brazil, v. 10, n. 2A (Suppl.), 2022. DOI: 10.15392/bjrs.v10i2A.1827. Disponível em: https://bjrs.org.br/revista/index.php/REVISTA/article/view/1827. Acesso em: 22 dec. 2024.

Similar Articles

1-10 of 332

You may also start an advanced similarity search for this article.

Most read articles by the same author(s)