Performance tests of a special ionization chamber for X-rays in mammography energy range

Authors

  • Jonas Oliveira da Silva Universidade Federal de Goiás
  • Linda V E Caldas Instituto de Pesquisas Energéticas e Nucleares - IPEN-CNEN/SP

DOI:

https://doi.org/10.15392/bjrs.v6i2A.501

Keywords:

ionization chamber, mammography, dosimetry

Abstract

A special mammography homemade ionization chamber was developed to be applied for mammography energy range dosimetry. This chamber has a total sensitive volume of 6 cm³ and is made of a PMMA body and graphite coated collecting electrode. Performance tests as saturation, ion collection efficiency, linearity of chamber response versus air kerma rate and energy dependence were determined. The results obtained with this special homemade ionization chamber are within the limits stated in international recommendations. This chamber can be used in quality control programs of mammography energy range. All measurements were carried out at the Calibration Laboratory of IPEN.

Downloads

Download data is not yet available.

Author Biographies

  • Jonas Oliveira da Silva, Universidade Federal de Goiás
    Instituto de Física
  • Linda V E Caldas, Instituto de Pesquisas Energéticas e Nucleares - IPEN-CNEN/SP
    Gerência de Metrologia das Radiações

References

IAEA - International Atomic Energy Agency. Absorbed dose determination in external beam radiotherapy: An international code of practice for dosimetry based on standards of absorbed dose to water. IAEA TRS 398, Vienna: IAEA, 2000.

IAEA - International Atomic Energy Agency. Dosimetry in diagnostic radiology: An international code of practice. IAEA TRS 457, Vienna: IAEA, 2007.

IAEA - International Atomic Energy Agency. Quality assurance programme for digital mammography. IAEA HHS 17, Vienna: IAEA, 2011.

ICRP - International Commission on Radiological Protection. Managing patient dose in digital radiology. ICRP 93, Ottawa: ICRP, 2004.

IEC - International Electrotechnical Commission. Medical electrical equipment—dosimeters with ionization chambers and/or semi-conductor detectors as used in X-ray diagnostic imaging. IEC 61674, Geneva: IEC, 1997.

IEC - International Electrotechnical Commission. Medical electrical equipment—Dosimeters with ionization chambers as used in radiotherapy. IEC 60731, Geneva: IEC, 2011.

MEGHZIFENE, A.; DANCE, D. R.; McLEAN, D.; KRAMER, H.-M. Dosimetry in diagnostic radiology. Eur. J. Radiol., v. 76, n.1, p. 11-14, 2010. DOI: https://doi.org/10.1016/j.ejrad.2010.06.032

NONATO, F. B. C.; SAKURABA, R. K.; CRUZ, J. C.; CALDAS, L. V. E. Characterization tests of a new parallel plate ionization chamber for use in electron beams. Rad. Phys. Chem., v. 104, p. 244-247, 2014. DOI: https://doi.org/10.1016/j.radphyschem.2014.06.011

PERINI, A. P.; NEVES, L. P.; VIVOLO, V.; XAVIER, M.; KHOURY, H. J.; CALDAS, L. V. E. Characterization of a CT ionization chamber for radiation field mapping. Appl. Radiat. Isot., v. 70, n. 7, p. 1300-1303, 2012. DOI: https://doi.org/10.1016/j.apradiso.2011.11.051

PTB - Physikalisch-Technische Bundesanstalt. Radiation qualities used for studies in radiation protection. Available at: . Last accessed: 31 Jan. 2018.

SHAFIRI, B.; ZEINALI, H. Z.; SOLTANI, J.; NEGARESTANI, A.; SHAHVAR, A. Investigation and performance tests of a new parallel plate ionization chamber with double sensitive volume for measuring diagnostic X-rays. Nucl. Inst. Meth. A, v. 770, p. 177-181, 2015. DOI: https://doi.org/10.1016/j.nima.2014.09.083

SILVA, J. O.; CALDAS, L. V. E. A double faced ionization chamber for quality control in diagnostic radiology beams. Appl. Radiat. Isot., v. 70, p. 1424-1428, 2012. DOI: https://doi.org/10.1016/j.apradiso.2011.11.052

YOSHIZUMI M. T.; CALDAS, L. V. E. Design and assembly of a simple monitor ionization chamber, In: Third European International Radiation Protection Association Congress, 2010, Helsinki. Proceedings. p. 14-16.

ZHU, T. C. Small Field: dosimetry in electron disequilibrium region. J. of Phys. Conf. Ser. v. 250, p. 1-10, 2010. DOI: https://doi.org/10.1088/1742-6596/250/1/012056

Downloads

Published

2018-07-19

How to Cite

Performance tests of a special ionization chamber for X-rays in mammography energy range. Brazilian Journal of Radiation Sciences, Rio de Janeiro, Brazil, v. 6, n. 2A (Suppl.), 2018. DOI: 10.15392/bjrs.v6i2A.501. Disponível em: https://bjrs.org.br/revista/index.php/REVISTA/article/view/501. Acesso em: 6 jan. 2025.

Similar Articles

1-10 of 248

You may also start an advanced similarity search for this article.

Most read articles by the same author(s)