Image quality evaluation for two different positron emitters in a preclinical PET scanner
DOI:
https://doi.org/10.15392/bjrs.v8i3.1299Palavras-chave:
image quality, preclinical PET scanner, two positron emitters, NEMA NU 4-2008Resumo
Positron emission tomography (PET) is widely used in preclinical trials, generating molecular images applied to biochemical, metabolic and functional investigation of organs and tissues. The positron emitters 11C and 18F radionuclides are relevant for different diseases studies. However, they have different positron energies, ranges, and branching ratio. This could result in a distinct quality between the acquired PET images. Thus, the aim of this study is to evaluate the image quality performance of the PET scanner (LabPET 4, GE) at CDTN/CNEN using the NEMA NU 4-2008 standards and specific phantom for these two different positron emitters. The NEMA image-quality (IQ) phantom consists of 3 different regions to analyze distinct characteristics: uniformity, recovery coefficients (RCs) and spill-over rations (SOR) in air and water. The IQ phantom was filled with two different aqueous solutions (18F-FDG and 11C-PK11195), both activities calibrated at the beginning of acquisition (3.7MBq). The IQ phantom was placed in the center of the field-of-view (FOV) and measured with the typical whole body imaging protocol. The images were reconstructed following the Molecular Imaging Laboratory (LIM/CDTN) standard protocol: MLEM-3D algorithm, 20 iterations, no high-resolution mode, no attenuation or scatter corrections, no post-filtering. PMOD® software was used to perform images post-processing following NEMA analyses standard. The %SD of medium concentration activity for 18F and 11C was 7.7 and 14.5, respectively. The RCs for 18F and 11C were 0.11 and 0.12 respectively for the 1-mm-diameter rod and 0.88 and 0.77 for the 5-mm-diameter rod. SORs in air were 0.19 and 0.22, respectively, and in water 0.29 and 0.33. Experimental results demonstrates that when 18F radionuclide is used, PET images presents a better performance in the image quality tests compared to 11C PET images.Downloads
Referências
R. Yao, R. Lecomte, E. Crawford. “Small-Animal PET: What is it, and why do we need it?” Journal of Nuclear Medicine Technology. Vol. 40 nº3, pp.157-165 (2011).
R. M. G. Gontijo. “Proposta de Programa de Garantia da Qualidade para Imagem Molecular Pré-Clínica”. 2019. 291 f. Thesis. (Doutorado em Ciência e Tecnologia das Radiações, Minerais e Materiais) – Centro de Desenvolvimento da Tecnologia Nuclear, Comissão Nacional de Energia Nuclear (CDTN/CNEN). Belo Horizonte. In Portuguese (2019).
National Electrical Manufacturers Association. “Performance Measurements of Small Animal Positron Emission Tomographs”. Rosslyn VA; 2008 Standards Publication NU 4-(2008).
IAEA - International Agency Energy Atomic, “Live Chart of Nuclides”. https://www-nds.iaea.org/relnsd/vcharthtml/VChartHTML.html (2019).
GE Healthcare Technologies, “Triumph Service Guide Technical Publication”. Revision Draft 6, Copyright. http://www.gehealthcare.com (2011).
M. A. Tretault, et al. “System architecture of the LabPET small animal PET scanner” IEEE Transactions on Nuclear Science, Vol.55, pp. 2546-2550, (2008).
R. Fontaine, et al. “The hardware and signal processing architecture of LabPETTM, a small animal APD-based digital PET scanner” IEEE Transactions on Nuclear Science, Vol. 56, pp. 3-9, February (2009).
Gamma Medica-Ideas, User Manual: Positron Emission Tomography (PET) Scanner Software - Manual for Acquisition and Processing of PET Images from the LABPET System. Software version: LABPET 1.12.0, (2010).
BELCARI, N. et al. NEMA NU-4 Performance Evaluation of the IRIS PET/CT Preclincial Scanner. IEEE Transactions on Nuclear Science, 2017. v.1, n.4, p. 301-309.
R. M. G. Gontijo. et al. “Image quality evaluation of a small animal PET scanner”Brazilian Journal of Radiation Sciences.Vol. 8 nº1, pp 01-13 (2020).
PMOD Tecnologies LCC, PMOD v.3.7. User Manual, (2015).
H. Zaidi and O. Ratib.“NEMA NU-04-based Performance Characteristics of the LabPET-8TM Small Animal PET Scanner” IEEE Nuclear Science Symposium Conference Record.pp. 3654-3660 (2011).
Downloads
Publicado
Edição
Seção
Licença
Direitos autorais (c) 2020 Brazilian Journal of Radiation Sciences

Este trabalho está licenciado sob uma licença Creative Commons Attribution 4.0 International License.
Declaro que o presente artigo é original, não tendo sido submetido à publicação em qualquer outro periódico nacional ou internacional, quer seja em parte ou em sua totalidade. Declaro, ainda, que uma vez publicado na revista Brazilian Journal of Radiation Sciences, editada pela Sociedade Brasileira de Proteção Radiológica, o mesmo jamais será submetido por mim ou por qualquer um dos demais co-autores a qualquer outro periódico. Através deste instrumento, em meu nome e em nome dos demais co-autores, porventura existentes, cedo os direitos autorais do referido artigo à Sociedade Brasileira de Proteção Radiológica, que está autorizada a publicá-lo em meio impresso, digital, ou outro existente, sem retribuição financeira para os autores.
Licença
Os artigos do BJRS são licenciados sob uma Creative Commons Atribuição 4.0 Licença Internacional, que permite o uso, compartilhamento, adaptação, distribuição e reprodução em qualquer meio ou formato, desde que você dê o devido crédito ao (s) autor (es) original (is) e à fonte, forneça um link para a licença Creative Commons, e indique se mudanças foram feitas. As imagens ou outro material de terceiros neste artigo estão incluídos na licença Creative Commons do artigo, a menos que indicado de outra forma em uma linha de crédito para o material. Se o material não estiver incluído no licença Creative Commons do artigo e seu uso pretendido não é permitido por regulamentação legal ou excede o uso permitido, você precisará obter permissão diretamente do detentor dos direitos autorais. Para visualizar uma cópia desta licença, visite http://creativecommons.org/licenses/by/4.0/